background image

DoUbLE RoLLER PUMP foR LAPARoScoPY 

  

   

Ackermann

Page: 16 // 18

13. APPENDIX

All product codes covered by these instructions are listed below:

• 

Double Roller Pump for Laparoscopy  

#16-2080 

14. ConTaCT DeTaIlS

ackermann Instrumente gmbH

Eisenbahnstrasse 65-67
78604 Rietheim-Weilheim
Germany

Phone:  +49 (0)7461 966 17 - 0
Fax: 

+49 (0)7461 966 17 - 70

E-Mail:   [email protected]
Web:   www.ackermannsurgical.com

End of document

Содержание 16-2080

Страница 1: ...SHAVER SYSTEM Ackermann www ackermannsurgical com Double Roller Pump for Laparoscopy 16 2080 User Manual Double Roller Pump ...

Страница 2: ...g set 8 6 5 Removing the tubing set 8 6 6 Using the device in operating room 8 7 Adapter duct pneumatic control 9 8 Setting up functional buttons control buttons and operation 10 8 1 Setting up 10 8 2 Functional test 10 8 3 Control buttons 11 8 3 1 AC Power switch 11 8 3 2 On off 11 8 3 3 Start stop 11 8 4 Operation of the device 11 9 Warning signals failure and fault description 12 10 Cleaning an...

Страница 3: ... IRRIGATION for pneumatic manual control or double footswitch 5 Connector SUCTION for pneumatic manual control or double footswitch 6 Display bargraph SUCTION 7 Roller wheel IRRIGATION 8 Roller wheel SUCTION 9 Tube holder 10 Fuse box 11 Mains power switch 12 Mains connection 13 Equipotentiality 14 Label next check 15 Model identification plate ...

Страница 4: ...tch Adapter duct with instrument connector autoclavable Certification CE0483 3 Indented use The device a suction irrigation pump without pressure monitoring and is used in minimal invasive surgery It can be utilized in all areas of endoscopic surgical operation procedures where the body cavities are not completely filled with fluid and where the volume of rinsing fluid is small in proportion to th...

Страница 5: ...ormation Name and address Model No Serial No Failure description Error message The unit consists of Pos Specification Article no Qty 1 Suction Irrigation pump 16 2080 1 2 Test report n a 1 3 User manual n a 1 4 Silicone suction irrigation tubing set autoclavable 16 2080 100 1 5 Main cable Euro 2 m 15 1720 800 1 6 Fuses 15 1720 901 2 7 Suction irrigation cannula dia 5 10 mm autoclavable 16 2080 201...

Страница 6: ...mains connection DANGER Do not open the unit No user servicable parts inside Symbol and description Symbol and description Symbol and description Follow the instructions for use Type BF applied part Year of production Legal manufacturer mdc medical device certification GmbH Kriegerstraße 6 70191 Stuttgart Germany Equipotential equalising potential Caution Hazardous voltage Batch number Article num...

Страница 7: ...segment Irrigation 19 Supply tube with connector for reservoir 20 Tube clamp 21 Spike connector 22 Adapter for instrument 23 Patient side tube section SUCTION for instrument 24 Pump segment SUCTION 25 Tube section for fluid collection receptacle Disassembled IMPORTANT The tube set may only be installed after the device has successfully performed a self test ...

Страница 8: ...ated Step 2 Take the red suction tube Place the tube stopper red into the upper tube holder marked red Proceed as described in step 1 Step 3 Next connect all connectors on the suction and irrigation tubes to the instrument to the reservoir and to the fluid collection receptacle 6 5 Removing the tubing set Detach all connectors between the instrument and reservoir before detaching the tube from the...

Страница 9: ...e irrigation and suction rates are proportional to the pressure applied to the buttons enabling a fine degree of control Pressing both buttons simultaneously has no effect Manual control with adapter duct A Suction irrigation cannula dia 5 mm B Suction irrigation cannula dia 10 mm C Blue button IRRIGATION D Suction irrigation duct E Red button SUCTION F Control connector IRRIGATION blue I SUCTION ...

Страница 10: ...t bargraph show the software version for 1 sec c 1st segment of the left bargraph display lights checking the pressure sensor for pneumatic control d 2nd 5th segments of the left bargraph light up checking the irrigation function of the motor e 6th 8th segments of the left bargraph light up checking the suction function of the motor f All segments are unlit again The device signals that is it is r...

Страница 11: ... after carrying out a functional check 2 The pump is activated the ON LED light green the STOP LED light yellow 3 Start the pump by pressing the push button START STOP The START LED light green 4 Pressing the red button on the manual control I right footswitch on the double footswitch starts the SUCTION roller wheel moving and pressing the blue button left footswitch the IRRIGATION roller wheel Th...

Страница 12: ...er 10 CLEANING MAINTENANCE To preserve the device and ensure its proper function please note as follows Cleaning of the device Please switch off the device after use Remove the power cable from the AC outlet Then remove the power cable from the AC connector on the rear of the device A disinfectant solution may be used to clean the exterior of the device Please consult the manufacturer s instructio...

Страница 13: ...of the suction and irrigation performance are valid as guidelines for the user inspection 12 quarterly inspections 12 1 checking the PNEUMATIC CONTROL MANUAL CONTROL DOUBLE FOOTSWITCH For this test you will need The device Pneumatic control manual control double footswitch The pneumatic controller test procedure is carried out in the following steps 1 Plug the AC plug into an AC outlet and switch ...

Страница 14: ...The irrigation procedure is activated 9 Pump fluid until fluid exits from the end of the tube in the measuring beaker A Continue to pump until there is approx 300 ml water in the measuring beaker Stop the irrigation procedure by pressuring the push button START STOP The STOP LED light yellow 1 0 Immerse the distal end of the cannula in the collected water in the measuring beaker A Press the red bu...

Страница 15: ...e Euro 2m 1 pc 15 17209 00 Fuses 10 pcs 16 2080 201 Suction Irrigation cannula dia 5 mm autoclavable 1 pc 16 2080 202 Suction Irrigation cannula dia 10 mm autoclavable 1 pc 16 2080 203 Adapter duct autoclavable 1 pc 16 2080 200 Pneumatic manual control autoclavable 1 pc 16 2080 204 Double lumen control tube autoclavable 1 pc 16 2080 1000 Pneumatic double footswitch 1 pc ...

Страница 16: ... instructions are listed below Double Roller Pump for Laparoscopy 16 2080 14 CONTACT DETAILS Ackermann Instrumente GmbH Eisenbahnstrasse 65 67 78604 Rietheim Weilheim Germany Phone 49 0 7461 966 17 0 Fax 49 0 7461 966 17 70 E Mail sales ackermannsurgical com Web www ackermannsurgical com End of document ...

Страница 17: ...Ackermann Page 17 18 SHAVER SYSTEM ...

Страница 18: ...ietheim Weilheim Germany User_Manual 16 2080 Revision_06 21 Copyright by Ackermann Instrumente GmbH Specifications design and accessories in this user manual are subject to change without any notice or obligation on the part of the manufacturer Ackermann 0483 ...

Отзывы: