![ACKERMANN 16-2050 Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/ackermann/16-2050/16-2050_user-manual_458095009.webp)
Ackermann
Page: 9 // 28
SHAVER SYSTEM
30. REv
: pedal button to switch the shaver blade in the Handpiece to anti-clockwise rotation. The number of rotations
depends on the set speed which can be adjusted via the Unit Control buttons (8) and (9) or Footswitch button (34).
31. OSCIllATION
: pedal button to switch to oscillating mode. The number of rotations depends on the set speed which
can be adjusted via the Control Unit buttons (8) and (9) or Footswitch button (34). The oscillation frequency depends on
settings.
32. FWD
: pedal button to switch the shaver blade in the Handpiece to clockwise rotation. The number of rotations depends
on the set speed which can be adjusted via the Control Unit buttons (8) and (9) or Footswitch button (34).
33. TOOL
: pedal button to initiate the change function of the steerable device “HANDPIECE 1” (2), “HANDPIECE 2” (3)
connected to the Control Unit.
34. SPEED
: pedal button to regulate the speed of the shaver blade in the Handpiece during use. Each time the button is
pressed, the speed is increased to the next higher level.
35. Handle
: used for lifting and carrying of the Footswitch.
36. Cable
: connection between the Footswitch and the Control Unit, serves to transmit the selected functions of the
individual Footswitch buttons (30-34).
37. Protective cap
: for increased stability of the connection and protection against fracture.
38. Plug
: connection to attach the Footswitch to the Control Unit. An integrated lock at the plug protects the cable of the
Footswitch from being pulled out by accident.
39. Receiver
: as a part of the Footswitch serves to transmit information from the Footswitch to the Control Unit, with
which is connected electrically.
40. Antenna
: necessary for communication between the Footswitch console and the receiver attached to the Control
Unit.
41. Charging socke
: used to connect the charger (45). Charging should be done outside of the treatment room. While
charging, functions of the Footswitch are disabled.
42. Connection lED
: flashing on blue diode signals the proper communication between the Footswitch console and the
receiver (39).
43. Battery charge status lED
: lighting on green, shows the accumulator level.
44. Error lED
: lighting on red diode signals the incorrect charger connection or not compatible charger connection.
45. Battery charger
: dedicated to charge the accumulator of the Footswitch. Should be used only outside the operating
room.
4. INSTAllATION AND STARTING ThE DEvICE
This chapter describes the correct installation and start-up of the device. The first step is the correct installation of the
Control Unit. The language select feature allows the operator to select the language of his choice. If the subsequent steps
described in chapters 4.1. to 4.6. have been followed correctly, the device is now ready for use and the shaver blades can
be attached.
Attention:
When the Shaver System Set is not in active use, protect the device against inadvertent actuation of the
Footswitch and the Handpiece!
4.1. DEvICE INSTAllATION
Before installation, make sure that the device has sufficient ventilation by maintaining a minimum distance of 10 cm from
the right, left and rear sides of the device.
Device installation:
• the only intended position of the Control Unit is horizontal, in which the device is placed on a flat surface on its four rubber
feet (12). In addition, adequate ventilation around the Control Unit must be provided,
• the place of installation should be a flat, dry and clean surface. This can be a table, shelf of an endoscopy trolley or other
elements meant for installation of medical devices.
Connecting the device to power:
• connect, using an appropriate lead, the equipotential bolt located on the rear of the Control Unit to an equipotential strip.
The cable insulation should be yellow-green,