SHAVER SYSTEM
1. PRECAUTIONS
IMPORTANT
• The operating instruction shows how to operate, maintain and service the equipment, so that its working life is as long as
possible. Proper service is essential for a long-term trouble-free operation of the device.
• Staff operating the device should familiarise themselves with the contents of this manual before installation and use of
the device, and those who have access to it.
• The device can only be operated by doctors or assisting staff with appropriate qualifications.
• The device is made up in I safety class.
•
CAUTION:
To avoid risk of electric shock the device has to be connected only to a supplying network with protective
earthing.
• Before connecting the device to the power outlet, make sure that the parameters of the power grid are identical to
those in the device specifications. Supply voltage, frequency and power consumption are specified in the remainder of
the documentation and on the rear panel of the device.
• Do not expose the device to strong humidity or rain.
•
WARNING!
Never use the device in the environments of flammable anesthetic gases.
• If flammable gases (including anesthetic gases) are released in the vicinity of the device, the latter should be immediately
switched off and unplugged from the power socket.
• To avoid overheating the device, ensure that adequate ventilation is provided before starting the device. It is recommended
to maintain a min. 10cm clearance from the left, right and rear sides of the device.
• Under no circumstances open the cover of the device when it is connected to power. Electric voltage inside the unit can
reach 0.4 kV. Electric shock may result in permanent disability or death.
• NEVER place the device near hot surfaces or areas subject to vibration or shock.
• Repairs of the device, other than those specified in this operating instruction, can be performed only by manufacturer
or by an authorised service provider specified by the manufacturer. The only repair that the user can perform, after
familiarising themselves with the contents of the operating instruction: changing fuses for the electric power connection.
Service address can be found in the last page of the operating instruction.
• Charging of the battery of wireless Footswitch should be carried out outside the operational theatre.
• Under no circumstances open the housing of Footswitch.
• Battery of wireless Footswitch mustn’t be replaced by the user.
• In case of any disturbances or malfunctions in the functioning of the device, immediately turn it off using the power
switch located on the rear panel. Taking into account the above guidance, the manufacturer feels responsible for
the safety and reliability of the device. The manufacturer does not assume responsibility for damages which have arisen
through abuse or incompatible with the intended operation of the device.