background image

Page 6

IV. NON-NORMAL OPERATION 

If either the batteries or the generating cell become depleted, or the 

generating cell is removed from the CAL-101, the instrument enters a non-

normal operating condition 

Under a low battery condition, the RED LEDs will begin to flash.  If the low 

battery condition occurs during power up, the RED LEDs will flash dimly  

for approximately 10 seconds and then the instrument will power down.  If 

the low battery condition occurs after start up, the  instrument will simply 

power down.  It is recommended to replace depleted batteries with two "AA" 

heavy duty Alkaline batteries for maximum operating life.

If the cell becomes depleted  either at start or during calibration, the GREEN 

LEDs will begin to 'ramp' up and down from max to min until the cell is 

removed. No gas generation will occur during this cell depleted indication. 

If there is an error in cell communication when inserted all green LED's 

will come on constant indicating a cell failure. Remove the cell and wait for 

all LED's to turn off then reinsert the cell ensuring proper alignment of the 

pins. If the error continues check the pins and corresponding pads on the 

cell for corrosion and attempt cleaning.

If the cell has been depleted, it should be removed and replaced with 

a new generating cell.  Please refer to the section on cell life for proper 

instructions on replacing the generating cells.  If the cell has not been 

depleted, there may be a faulty electrical connection between the cell and 

the instrument.  Remove the cell and inspect the pins protruding from the 

pump board and the corresponding pads on the cell board. Remove any 

debris and or corrosion that may be present and reinsert the cell ensuring 

proper alignment.  If either the pins  become damaged or non-repairable, 

the pump board should be replaced..

The  rate  of  airflow  through  the  instrument  is  critical  for  accurate  gas 

concentrations. The flow should be adjusted to 0.5 LPM (liters per minute) 

STP (standard temperature and pressure) of sea level and 20

o

C. (see chart 

in section VI) To trim the flow hold the magnetic screw driver over the ‘flow 

trim’ area of the instrument indicated with a magnet symbol. The GREEN 

LEDs will begin flashing at a once per second rate indicating the relative 

power to the pump. As you hold the magnet over this area the pump will 

continue to increase in speed (The LEDs will ramp up indicating the relative 

power) until the magnet is removed. When the magnet is reapplied

Содержание CAL-101

Страница 1: ...InstructionManual CalibrationGasGenerator AdvancedCalibrationDesigns Inc CAL 101...

Страница 2: ...al Description Page 2 II Start Up Page 4 III Normal Operation Page 5 IV Non Normal Operation Page 6 V Cell Life Page 8 VI Corrections to Volumetric Flow Page 9 VII Maintenance Page 10 VIII Accessory I...

Страница 3: ...can lead to improper calibrations This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Stat...

Страница 4: ...is produced when the CAL 101 supplies a known amount of current and a known amount of air continuously to the cell Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides 10 12...

Страница 5: ...ed The bump generating cell is equipped with a switch to initiate a bump cycle In this configuration the cell remains in the instrument and automatically turns off after a 30 second bump cycle Deliver...

Страница 6: ...ndicates that 2 hours remain on the cell 1 LED lit indicates less than 12 minutes of calibration time remain These are the normal operating conditions for the CAL 101 Complete equilibrium may take an...

Страница 7: ...the delivery hose This adapter or cup is then placed over the gas sensor Be sure not to restrict or block the delivery hose while the pump is running This method is used for calibrating diffusion typ...

Страница 8: ...corresponding pads on the cell for corrosion and attempt cleaning If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life fo...

Страница 9: ...er minute STP standard temperature and pressure of sea level and 20o C see chart in section VI To trim the flow hold the magnetic screw driver over the flow trim area of the instrument indicated with...

Страница 10: ...plete list of correct cell part numbers please refer to the parts list at the end of this manual To replace the generating cell in the instrument press down firmly on the cell eject button at the top...

Страница 11: ...ilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Temperature Correction Table Altitude Correc...

Страница 12: ...t fitting body panel mount 400 0000 00 Circuit board for CAL 101 Processor 400 0001 00 Circuit board for CAL 101 Interface 400 0002 00 Circuit board for CAL 101 pump assy 510 0101 00 Chlorine generati...

Страница 13: ...ation Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing...

Страница 14: ...tured by Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Web site www GoACD com E mai...

Страница 15: ...2024W McMillanSt Tucson Arizona85705USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0601 00 AdvancedCalibrationDesigns Inc...

Отзывы: