III. NORMAL OPERATION

 

After the desired equilibrium has been achieved, the instrument may be 

used to perform gas calibration.  The GREEN LEDs will continue to flash 

ever 4 seconds indicating remaining cell life.  While in normal operating 

mode, the CAL-101 will run continuously and generate gas until either the 

battery becomes discharged or the cell is consumed. (In the bump station 

configuration the unit will run for thirty seconds then automatically turn off)  

If the cell is consumed the green LEDs will 'ramp' max to min repeatedly 

indicating the need for cell replacement. When this occurs, please refer to 

the Non-Normal Operation section that follows.

Note: Changes in air pressure or temperature will affect the output of 

the CAL-101 instrument.  Before beginning any calibration please refer 

to section VI. CORRECTIONS TO VOLUMETRIC FLOW- for important 

adjustments needed to obtain accurate calibration gas.

The calibration gas is delivered through the calibration hose attached to the 

CAL-101 via a quick connect coupling.  A calibration adapter, or cup, which 

is provided by the manufacturer of your gas detector, is usually attached 

to the end of the delivery hose.  This adapter or cup is then placed over 

the gas sensor.  Be sure not to restrict or block the delivery hose while the 

pump is running.  This method is used for calibrating diffusion type sensors.

Alternatively, the output gas may be collected into a gas collection bag 

for calibrating sample draw instruments.  If this method is used, it is 

recommended that the instrument is allowed to flow into and out of the bag 

via a “Y” connector.  First, partially fill the bag with generated gas.  Second, 

attach the sample draw unit to the third side of the “Y” connector.  The 

sample draw instrument will draw from both the generator and/or the bag 

depending upon the flow rate of the generator.  Be sure to disconnect the 

sample draw unit before the bag is completely emptied of gas.

An alternative method is to connect the “Y” adapter directly to the instrument 

with no bag attached.  This should only be used if the flow rate of the CAL-

101 is greater than that of the sample draw instrument, otherwise dilution 

gas will also be drawn into the instrument reducing the calibration gas 

concentration.  

NOTE: Never attach a sample draw instrument directly to the end of 

the CAL-101 delivery hose without a “Y” connector in between the two 

instruments.  This may cause damage to one or both of the air pumps 

in the CAL-101 and/or sample draw instrument.

 

Page 5

Содержание CAL-101

Страница 1: ...InstructionManual CalibrationGasGenerator AdvancedCalibrationDesigns Inc CAL 101...

Страница 2: ...al Description Page 2 II Start Up Page 4 III Normal Operation Page 5 IV Non Normal Operation Page 6 V Cell Life Page 8 VI Corrections to Volumetric Flow Page 9 VII Maintenance Page 10 VIII Accessory I...

Страница 3: ...can lead to improper calibrations This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Stat...

Страница 4: ...is produced when the CAL 101 supplies a known amount of current and a known amount of air continuously to the cell Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides 10 12...

Страница 5: ...ed The bump generating cell is equipped with a switch to initiate a bump cycle In this configuration the cell remains in the instrument and automatically turns off after a 30 second bump cycle Deliver...

Страница 6: ...ndicates that 2 hours remain on the cell 1 LED lit indicates less than 12 minutes of calibration time remain These are the normal operating conditions for the CAL 101 Complete equilibrium may take an...

Страница 7: ...the delivery hose This adapter or cup is then placed over the gas sensor Be sure not to restrict or block the delivery hose while the pump is running This method is used for calibrating diffusion typ...

Страница 8: ...corresponding pads on the cell for corrosion and attempt cleaning If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life fo...

Страница 9: ...er minute STP standard temperature and pressure of sea level and 20o C see chart in section VI To trim the flow hold the magnetic screw driver over the flow trim area of the instrument indicated with...

Страница 10: ...plete list of correct cell part numbers please refer to the parts list at the end of this manual To replace the generating cell in the instrument press down firmly on the cell eject button at the top...

Страница 11: ...ilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Temperature Correction Table Altitude Correc...

Страница 12: ...t fitting body panel mount 400 0000 00 Circuit board for CAL 101 Processor 400 0001 00 Circuit board for CAL 101 Interface 400 0002 00 Circuit board for CAL 101 pump assy 510 0101 00 Chlorine generati...

Страница 13: ...ation Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing...

Страница 14: ...tured by Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Web site www GoACD com E mai...

Страница 15: ...2024W McMillanSt Tucson Arizona85705USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0601 00 AdvancedCalibrationDesigns Inc...

Отзывы: