I. GENERAL DESCRIPTION

The CAL-101 is a small, battery-powered, portable electrochemical gas 

generator designed to calibrate toxic gas sensors.  

Fast warm-up time allows the instrument to be turned 

off between remotely located sensors saving battery 

life and avoiding generation of unwanted gas.  The 

CAL-101 uses the following components to produce 

the calibration gas/air mixture:

Internal Micro Pump

A small air pump draws in ambient air to blend with the electrochemically 

generated gas.

Electrochemical Generating Cell

The electrochemical generating cell contains an electrolyte solution and 

either inert or consumable electrodes, depending upon the gas being 

generated.  A precise concentration of gas is produced when the CAL-101 

supplies a known amount of current and a known amount of air continuously 

to the cell.  

Alkaline “AA”  Batteries

A set of two fully charged alkaline “AA” batteries provides 10-12 hours of 

operation.  To replace the batteries, first remove the cell then slide the battery 

chamber in to allow the batteries to drop out the bottom of the instrument.  

NOTE:   Rechargeable batteries will provide substantially fewer hours 

of operation and will self discharge while the instrument is in storage. 

Users should also be aware that use of rechargeable batteries will void 

the CSA approval of the CAL-101 for intrinsic safety.

Microprocessor-Based Circuitry 

When powered-up, the microprocessor provides remaining cell  and battery 

life indication via the display LEDs. Low battery or exhausted cell conditions 

are also indicated when the condition occurs.  (See section IV Non-Normal 

Operation.)  

Page 2

Содержание CAL-101

Страница 1: ...InstructionManual CalibrationGasGenerator AdvancedCalibrationDesigns Inc CAL 101...

Страница 2: ...al Description Page 2 II Start Up Page 4 III Normal Operation Page 5 IV Non Normal Operation Page 6 V Cell Life Page 8 VI Corrections to Volumetric Flow Page 9 VII Maintenance Page 10 VIII Accessory I...

Страница 3: ...can lead to improper calibrations This instrument conforms to the protection requirements of the EC DIRECTIVE 89 336 EEC on Electromagnetic Compatibility EMC in accordance with the provisions of Stat...

Страница 4: ...is produced when the CAL 101 supplies a known amount of current and a known amount of air continuously to the cell Alkaline AA Batteries A set of two fully charged alkaline AA batteries provides 10 12...

Страница 5: ...ed The bump generating cell is equipped with a switch to initiate a bump cycle In this configuration the cell remains in the instrument and automatically turns off after a 30 second bump cycle Deliver...

Страница 6: ...ndicates that 2 hours remain on the cell 1 LED lit indicates less than 12 minutes of calibration time remain These are the normal operating conditions for the CAL 101 Complete equilibrium may take an...

Страница 7: ...the delivery hose This adapter or cup is then placed over the gas sensor Be sure not to restrict or block the delivery hose while the pump is running This method is used for calibrating diffusion typ...

Страница 8: ...corresponding pads on the cell for corrosion and attempt cleaning If the cell has been depleted it should be removed and replaced with a new generating cell Please refer to the section on cell life fo...

Страница 9: ...er minute STP standard temperature and pressure of sea level and 20o C see chart in section VI To trim the flow hold the magnetic screw driver over the flow trim area of the instrument indicated with...

Страница 10: ...plete list of correct cell part numbers please refer to the parts list at the end of this manual To replace the generating cell in the instrument press down firmly on the cell eject button at the top...

Страница 11: ...ilized Note Altitude and Temperature are cumulative i e you must correct for both factors to obtain the most accurate calibration Altitude Correction Table Temperature Correction Table Altitude Correc...

Страница 12: ...t fitting body panel mount 400 0000 00 Circuit board for CAL 101 Processor 400 0001 00 Circuit board for CAL 101 Interface 400 0002 00 Circuit board for CAL 101 pump assy 510 0101 00 Chlorine generati...

Страница 13: ...ation Designs Inc s obligation under this warranty shall be limited to repairing or replacing and returning any product which shall be returned to Advanced Calibration Designs Inc at its manufacturing...

Страница 14: ...tured by Advanced Calibration Designs Inc 2024 W McMillan St Tucson Arizona 85705 USA Phone 520 290 2855 Fax 520 290 2860 Int l Phone 001 520 290 2855 Fax 001 520 290 2860 Web site www GoACD com E mai...

Страница 15: ...2024W McMillanSt Tucson Arizona85705USA Telephone 520 290 2855 Fax 520 290 2860 TollFree 800 737 0223 Email acd goACD com Web www goACD com P N 910 0601 00 AdvancedCalibrationDesigns Inc...

Отзывы: