Control requirement
:
According to high temperature temperature SV and row of wind temperature SV
the corresponding relationships, decided when [EXHAUST] does contact
outputs. When high-temperature test execution, take high-temperature test
temperature SV as the determination object; When preheating experimental
execution, take preheating temperature SV as the determination object.
Acts the condition
HIGH MODE
Controlled member temperature > row of wind hypothesis temperature time,
outputs the OFF.
Controlled member temperature
≦
row of wind hypothesis temperature time,
outputs ON.
LOW MODE
Controlled member temperature > row of wind hypothesis temperature time,
outputs the ON.
Controlled member temperature
≦
row of wind hypothesis temperature time,
outputs OFF.
13. Initial period setup 4
13.1
除霜待機
-Defrosts the waiting
“Defrosts automatically the setup” the choice “The time” time, this function is only then effective.
1.
有
-enable: After now experimented the cold temperature test finally, started to carry out again
defrosts the movement.
2.
無
-Disable: The setup time arrived, the cold temperature test interrupts temporarily immediately,
and starts to carry out defrosts the movement; The waiting defrosted finally, and after
the test condition satisfied, then continued to complete has not carried out the cold
temperature test.
NOTE:ALALM screen in the screen bottom, Only appear when you have already setup in output
relay number.
52