10.3
線性回歸
- Linear adjust
This function supplements and corrects for temperature SENSOR, may revise gets older SENSOR the
error or revises the non-linear curve for the linear curve; The preheating area temperature, the
precooling area temperature, the test area temperature, the freezer temperature have 5 adjusting
points individually.
Setup scope
: -99.99 ~ 325.00˚C
Corrected values and hope value relations:
-210˚C< set value 1
≦
set value 2
≦
set value 3
≦
set value 4
≦
set value 5
≦
325˚C
Revision case 1
: Hope controller present demonstration value 50.00˚C revision for demonstration
48.00˚C; So long as then the adjusting point hypothesis for 50.00˚C, hoped that an hypothesis for
48.00˚C, then changes the demonstration temperature.
Revision case 2
: We discovered that when sends in the standard signal 0˚C controller
demonstration 2˚C, when sends in the standard signal 100˚C controller demonstration for 98˚C,
deviation of the controller obviously for non-grade condition; This time like must revise this error only
to in the temperature column, adjusting point “000.00” the position enters “002.00”; Adjusting point
“100.00” the position enters “098.00” then with ease to complete the revision.
27