11.1 N2GAS
延遲時間設定
- N2GAS delay time setup
Setup scope: 0 ~ 99 minutes
11.1.1 Acts the condition
The high-temperature test or the normal temperature experiment ended, carried out toward the
cold temperature test, “N2GAS” the output was only then effective.
1. transforms for the cold temperature test, the controller starts the delay time which calculates
establishes; After the time arrives, if the test area temperature's actual value has not achieved the
cold temperature test the setting value (note one), N2GAS output “ON”.
2. The test area temperature's actual value achieves the cold temperature test setting value time (note
one), N2GAS output “OFF”.
3. Outside the above two control requirement, the N2GAS output has maintained at “OFF” condition.
Note 1 : If has the hypothesis “to wait for an opportunity” the temperature, must wait for an
opportunity the condition to relieve is only then effective.
11.1.2 Acts the characteristic chart
ON
N2GAS output
Cold temperature
N2GAS delay time
High-temperature test
or normal temperature
experiment
PV
11.2
防汗開始溫度
- Against perspiration initial temperature setup
Setup scope: - 9 ~ + 9˚C.
30