ProPanel-HD (FW4), Revised August 2020
Page 11 of 24
Edge to Hole Measurements
1. Close the jaws and push the DATUM key.
2. Install a full cone onto the moving jaw, and the half cylinder onto the fixed
jaw. For the most accurate measurements, be sure the flat edge of the half
cylinder is installed co-planer to the flat edge of the measurement jaw.
3. Press the F1 key until the #3 superscript appears on the screen.
Notes:
•
The tips of the inserts are even with the edges of the jaws, so the
value displayed is correct.
•
Edge-to-
hole measurements less than 0.75” are not possible.
4. Open the jaws and place the fixed jaw with the half cylinder against the
outside edge of the part. Then place the full cone into the hole.
5. Read the Hole-to-Edge distance on the readout.
6. If you need to recalibrate with these inserts installed:
a. Press the DATUM key to clear the readout.
b. Close the moving jaw so the cones touch.
c. Press the F1 key until the #3 superscript appears.
d. The reading should equal 0.750 inches.
Note: Videos showing how these measurements are made are posted at web
site: