Pr12
– Readout Auto-Off Time
[0 to 240]
This parameter changes the amount of time before the ProPanel Readout
turns off to conserve battery power. The value is the number of minutes of
idle operation (no movement or key presses) before the display turns off.
When this parameter is enabled (
0
), pressing the
ON
/
OFF
key or movement
of the system will wake up the Readout.
A valu
e of ‘
0
’ disables the Auto Off feature (the Readout is always on).
Default Value = 15 (minutes)
Pr13
– Linear Compensation
[0.00001 to 9.99999]
This parameter invokes a linear multiplier correction in the Readout that is
applied to the actual measurement prior to any offsets. This adjusted
measurement is then displayed on the Readout. This is useful, for instance, if
you would like to display half, double, or other values that can be achieved by
multiplying the actual measurement by the value of the parameter setting.
SEE PAGE 2 FOR THE CUSTOM VALUE FOR THIS PROPANEL
Pr14
– ProScale Compatibility
[0 or 1]
This parameter sets the Readout to operate with the correct technology.
This parameter is preset at the factory if the Readout is purchased as part of
a complete System. See also Segment Offset.
FACTORY SET FOR PROPANEL TO 0
Pr15-Pr21 NOT USED
Pr22
– ABS/INC Key Operation
[0 or 1]
This parameter controls the amount of time the
ABS
/
INC
key needs to be
pressed to enter the INCremental measuring mode.
Default Value = 0
Pr23
– Auxiliary Keypad Enable
[0 to 3]
The parameter controls the operation of the
ABS
/
INC
and
SEND
keys.
Default Value = 3
Pr 22 Setting
Key press to enter INCremental mode
0
Press and Hold
ABS/INC key (1.2 sec)
1
Momentary
depress ABS/INC key (< 1 sec)
Pr 23 Setting
ABS/INC and SEND keys
0
Disable Both Keys
1
Enable abs/inc Key only
2
Enable send Key only
3
Enable Both Keys