General Maintenance
ProPanel will operate in a dry environment with non-conductive debris such
as sawdust, plastic, dust, and small amounts of water splash with no adverse
effects. The system should be cleaned of excess debris when necessary to
prevent premature damage to the Scale or Encoder.
The Digital Readout should be cleaned periodically with compressed air to
remove any dust on the lens and keys.
All mounting fasteners should be checked occasionally for tightness.
Occasionally check parallelism of the jaws by measuring a piece of paper
between the upper tips of the jaws, then between the lower tips of the jaws. A
variation of 0.05mm (0.002 inches) is within specifications.
If the jaw becomes difficult to move, verify the scale is clean. Find and
remove any burrs which may have developed on the aluminum scale.
Do not use any liquid lubricants on the ProPanel system.
Recalibration or Repairs may be arranged with Accurate Technology by
completing the Return Material Authorization forms located at
proscale.com>Customer Support>Repair Request.
Battery Replacement
When the Readout battery needs to be replaced, a low battery indicator will
appear in the lower left corner of the LCD. Press and hold the
DATUM
key for
6 seconds to display the current battery voltage.
When battery voltage drops below 2.6v, the Readout will turn itself off until the
battery is replaced
.
Remove the screws in the upper right and lower left corners of the Readout.
Pull the cover off. Remove the old battery. Install a new CR123 (or
equivalent) battery noting the proper orientation. Replace the cover and
screws.
–
+