AM8060 Precision Thermometer User’s Guide
5/41
2
Specifications
2.1 Specifications
Temperature range
:
-200 ºC ~ 850 ºC
Resolution
:
0.001 ºC
Accuracy
:
25Ω SPRT/IPRT
100Ω SPRT/IPRT
±0.01 ºC @ -196 ºC
±0.010 ºC @ -200 ºC
±0.008 ºC @ 0 ºC
±0.008 ºC @ 0 ºC
±0.009 ºC @ 232 ºC
±0.009 ºC @ 232 ºC
±0.015 ºC @ 420 ºC
±0.015 ºC @ 420 ºC
±0.025 ºC @ 660 ºC
±0.020 ºC @ 660 ºC
±0.025 ºC @ 850 ºC
Short term stability
:
o
±0.001 ºC
(
with 100Ω platinum probe
)
o
±0.003 ºC
(
with 25Ω platinum probe
)
Working environment
:
15ºC ~ 35 ºC
Power supply
:
AC 100V ~ 240V
Power consumption
:
< 15 W
Dimensions
:
180 mm x 65 mm x 200 mm
Weight
:
0.7 lbs (300 g)
2.2 Probe Types
SPRT or PRT with temperature coefficient α=0.003925
(W
Ga
≥1.11807).
IPRT with temperature coefficient α=0.00385.