AM8060 Precision Thermometer User’s Guide
21/41
(underscore) to the desired, and using
or
key to adjust the
value. After the adjustment, the instrument will display
:
24.9977
Rt =24.9977 OHM
Menu= Exit Enter = Yes
25R Rs = OHM
* 25 Ohm CALIBRATION
5.2.7
It is required to wait for 10 seconds to ensure the reading is
stable. Each second is shown with a blinking dot at the top left
hand corner of the display. The next step is to save the calibrated
data to the instrument. To start the data saving, touch the
MENU
key, and the third line of data will blink to indicate the option to
save.
5.2.8
To save this calibration, press and hold the
key and then
press the
key. This will save the calibration data to the
instrument, and the program will go back to the main menu
(*CALIBRATION MENU*)
;
If users DO NOT want to save
the data, press the
MENU
key will bring the program back to the
main menu ( *CALIBRATION MENU*).
5.3
Reset the measurement/calibration switch after calibration
5.3.1
After the calibration, turn off the AC power
5.3.2
Set the Measurement/Calibration switch at the rear panel to the
left side (away from “CAL”)