EXI-600 INVERTED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
20
5.2.13.
Fluorescence sliders (optional; fluorescence only)
Fully insert the Reflected light field
aperture slider into the “F” slot of
the mainframe. Note the
orientation of the white triangle
indicator for light throughput.
Detents in the slider help identify
the “out” and “in” positions.
Fully insert the Reflected light
aperture diaphragm slider
(indicated with an “A”) into the “A”
slot of the mainframe. Detents in
the slider help identify the “out”
and “in” positions.
Insert the 3-position Neutral
Density (“ND”) filter slider into the
rear-most slot (to the right of the
Aperture iris slider).
Starting adjustment for
fluorescence: Set the Field and
Aperture sliders to the top (fully
open) positions. Using the detents
on the ND filter slider, push in or
pull out the slider until the middle
(empty) position is in the light path
– this position has no attenuation
of fluorescence signal and is a
good starting point for
fluorescence observation.
Reflected light
field diaphragm
slider
Blank
slider
Reflected
light
aperture
diaphragm
slider
Neutral density
filter slider