EXI-600 INVERTED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
11
5.2.2.
Eyepieces
Remove the dust plugs from the
eyetubes.
Insert the eyepieces into the
eyetubes.
5.2.3.
Transmitted illumination pillar
Use a 4mm hex screwdriver and
remove the 4 hex screws from the
back of the mainframe. These
screws will be used again
momentarily.
Hold the pillar above the
mainframe with the base of the
pillar directly above the mounting
position.
Align the guide pins with the slots
in the frame.
Slowly slide the pillar down into the
slots. The pillar is not yet secured,
but will not fall over.
Secure the pillar onto the frame
using the same 4 hex screws from
above.
When installed properly, the pillar
can be easily tilted backward and
forward with one hand.
Dust plug
Eyepiece
Transmitted
illumination
pillar
4mm hex screws