EXI-600 INVERTED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
45
used
for
the
observation
technique from the light path.
Light flickers.
Poor electrical connection.
Check all power cables to the
LED lamp house and microscope
are firmly attached.
OBSERVATION TUBE
Problem
Cause
Corrective Measure
The view with one eye does not
correlate with the view with the
other eye.
Interpupillary distance between
eyepieces is incorrect.
Adjust the interpupillary distance
of the binocular eyetubes until a
single, overlapping image is
observed when viewing with both
eyepieces simultaneously.
Eyepiece diopter is not properly
adjusted.
Adjust the diopter of one
eyepiece so that the focus of the
image
matches
that
seen
through the other eyepiece.
Still adapting to viewing through
the
eyepieces
of
this
microscope.
It takes time to adjust to a
different microscope. Try looking
around the periphery of the field
of view before concentrating on
the specimen. Looking far away
from the microscope (e.g., at an
object across the room) before
and during microscope use can
aid eyes in acclimating to a
microscope (and reduce fatigue).
10.3.
MICROSCOPY IMAGING
Problem
Cause
Corrective Measure
Image not in focus.
Edge or periphery of image is
blurry compared to rest.
Ghost image of window, interior
room light, etc. appears in
camera image.
Light entering eyepieces and
merging with specimen image.
Cover eyepieces when imaging.
Move camera to different camera
port that allows 100% of light to
go to the camera.