ASSEMBLY 

Condenser 

To install the condenser: 

1.  Unscrew the condenser set screw sufficiently to allow 

the condenser tube to slide over the dovetail groove of 

the condenser hanger. 

2.  Lightly press the condenser into position and tighten 

the set screw. 

Phase Contrast Slider 

To install the phase contrast slider: 

1.  With the printed notations on the slider facing up and 

readable from the front of the microscope, insert the 

phase contrast slider horizontally into the condenser 

slot.  The orientation of the slider is correct if the edge 

of the slider facing the operator has adjusting screws 

visible. 

2.  Continue to insert the slider until an audible "click" 

indicates that one position of the 3-postion phase 

contrast slider is aligned with the optical axis.  Insert 

the slider further into the slot or backward to the 

desired slider position.  

Mechanical Stage (optional) 

To install the optional mechanical stage:

 

 

1.  Install the mechanical according to path 

 (as shown 

in the figure).  First, align the edge A of the mechanical 
stage with the edge of the flat/plain stage surface.  

Align the mechanical stage with the plain stage until 

the two set screws in the bottom of the mechanical 

stage align with screw holes in the bottom of the plain 

stage.  Tighten the two set screws. 

2.  Install the universal holder according to path 

 (as 

shown in the figure).  Start by placing the flat universal 

holder plate on the plain stage surface.  Align the two 

screw holes on the universal holder plate with the set 

screws on the lateral movement ruler of the mechanical 

stage.  Tighten the two set screws. 

Set screw 

Adjustment screws 

Содержание EXI-410 Series

Страница 1: ...MANUAL 73 Mall Drive Commack NY 11725 631 864 1000 P 631 543 8900 F www accu scope com info accu scope com EXI 410 INVERTED MICROSCOPE SERIES...

Страница 2: ......

Страница 3: ...OPERATION POWERING ON THE MICROSCOPE 18 ILLUMINATION 18 INTERPUPILLARY DISTANCE 18 FOCUSING 19 ADJUSTABLE TENSION CONTROL 20 STAGE PLATES OPTIONAL 20 SELECTING THE LIGHT PATH 21 APERTURE DIAPHRAGM ADJ...

Страница 4: ...void blocking the natural air circulation for cooling Ensure objects and obstructions are at least 10 centimeters from all sides of the microscope the only exception is the table on which the microsco...

Страница 5: ...should be swabbed using a circular motion A small amount of absorbent cotton wound on the end of a tapered stick such as cotton swabs or Q tips makes a useful tool for cleaning recessed optical surfac...

Страница 6: ...roscope if needed Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form Carefully remove the microscope from the EPE foam container by i...

Страница 7: ...Emboss Contrast Slider 6 Power Indicator 7 Illumination Selector 8 Main Frame 9 LED Lamp transmitted 10 Illumination Pillar 11 Condenser Set Screw 12 Field Iris Diaphragm 13 Condenser 14 Objective 15...

Страница 8: ...illar 2 Field Iris Diaphragm 3 Phase Contrast Slider 4 Condenser 5 Mechanical Stage with Universal Holder optional 6 Objective 7 Nosepiece 8 Power Switch 9 Eyepiece 10 Eyetube 11 Viewing Head 12 Light...

Страница 9: ...rast 1 Viewing Head 2 Stage 3 Emboss Contrast Slider 4 Main Frame 5 Focus Tension Adjustment Collar 6 Coarse Focus 7 Fine Focus 8 Condenser Set Screw 9 Phase Contrast Slider 10 Condenser 11 Illuminati...

Страница 10: ...elector 9 Main Frame 10 LED Lamp transmitted 11 Illumination Pillar 12 Condenser Set Screw 13 Field Iris Diaphragm 14 Condenser Centering Screw 15 Condenser 16 Light Shield 17 Objective 18 Stage 19 Me...

Страница 11: ...ider 4 Condenser 5 Mechanical Stage with Universal Holder optional 6 Objective 7 Nosepiece 8 Fluorescence Turret 9 Fluorescence Turret Access Door 10 Power Switch 11 Eyepiece 12 Eyetube 13 Viewing Hea...

Страница 12: ...ld 3 Emboss Contrast Slider 4 Main Frame 5 Focus Tension Adjustment Collar 6 Coarse Focus 7 Fine Focus 8 Condenser Set Screw 9 Phase Contrast Slider 10 Condenser 11 Illumination Pillar 12 Light Shield...

Страница 13: ...410 INVERTED MICROSCOPE SERIES ACCU SCOPE 73 Mall Drive Commack NY 11725 631 864 1000 www accu scope com 12 MICROSCOPE DIMENSIONS EXI 410 Phase Contrast and Brightfield EXI 410 FL with Mechanical Sta...

Страница 14: ...ble the various components The numbers indicate the order of assembly Use the hex wrenches that are supplied with your microscope when required Be sure to keep these wrenches for changing out componen...

Страница 15: ...ible click indicates that one position of the 3 postion phase contrast slider is aligned with the optical axis Insert the slider further into the slot or backward to the desired slider position Mechan...

Страница 16: ...tive onto the nosepiece opening then rotate the nosepiece clockwise and thread the other objectives from low to high magnification NOTE Always rotate the nosepiece by using the knurled nosepiece ring...

Страница 17: ...ficant vignetting c A 0 5X lens optional accommodates camera sensors and smaller Larger sensors may result in images with significant vignetting Fluorescence Filter Cubes EXI 410 FL models only To ins...

Страница 18: ...than indicated will cause severe damage to your microscope Connecting the Power Cord Make sure the On Off Switch is O the off position before connecting the power cord Insert the power plug into the...

Страница 19: ...g upon the specimen density and objective magnification Adjust the light intensity for comfortable viewing by turning the light intensity control knob clockwise towards the operator to increase bright...

Страница 20: ...e sharp image using your right eye do not touch the coarse or fine adjustments Instead rotate the right diopter collar until the sharpest image appears Repeat several times to check IMPORTANT do not c...

Страница 21: ...tes Optional NOTE for optimal viewing ensure the thickness of the container dish or slide matches the thickness marked on each objective 0 17mm or 1 2mm For modern objectives coverglass is optimally 0...

Страница 22: ...ewing head IN position or 100 to the camera port OUT position Using the Aperture Diaphragm The iris diaphragm determines the numerical aperture N A of the illumination system in bright field observati...

Страница 23: ...pically necessary If the phase ring is not centered you can adjust it by centering the bolt with the 2mm hex wrench provided with the microscope see instructions below The EXI 410 PH includes a 3 posi...

Страница 24: ...nulus may sometimes appear It this occurs superimpose the brightest light annulus image over the phase annulus When a thick specimen is moved or replaced the light annulus and the phase annulus may de...

Страница 25: ...s Contrast Slider The eyepiece tube side emboss contrast slider has several position markings corresponding to the objective magnification and several stop positions to ensure alignment of the apertur...

Страница 26: ...able with alternate light sources for fluorescence Various filter sets are also available for installation Selection of filter sets depends upon the available LED fluorescence modules in your microsco...

Страница 27: ...d the fluorescence LED unit is also automatically switched Brightfield positions on the turret are indicated by a symbol and alternate with the three fluorescence filter cube positions Detents on the...

Страница 28: ...r position Move the slider until it clicks into place Dirt or dust is visible in the field of view Dirt dust on the specimen Clean or replace the specimen or Dirt dust on the eyepiece Clean the eyepie...

Страница 29: ...ate is poor profile irregularity etc Use a vessel with a good profile irregularity characteristic MECHANICAL PART PROBLEM CAUSE SOLUTION The coarse adjustment knob is too difficult to rotate The tensi...

Страница 30: ...Adjust the diopter Your view is not accustomed to the microscope observation and widefield eyepieces Upon looking into eyepieces try looking at the overall field before concentrating on the specimen...

Страница 31: ...ce to the original end user purchaser LED lamps are warranted for a period of two years from the date of original invoice to the original end user purchaser The mercury power supply is warranted for a...

Отзывы: