EXI-410 INVERTED MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
14
ASSEMBLY
Condenser
To install the condenser:
1. Unscrew the condenser set screw sufficiently to allow
the condenser tube to slide over the dovetail groove of
the condenser hanger.
2. Lightly press the condenser into position and tighten
the set screw.
Phase Contrast Slider
To install the phase contrast slider:
1. With the printed notations on the slider facing up and
readable from the front of the microscope, insert the
phase contrast slider horizontally into the condenser
slot. The orientation of the slider is correct if the edge
of the slider facing the operator has adjusting screws
visible.
2. Continue to insert the slider until an audible "click"
indicates that one position of the 3-postion phase
contrast slider is aligned with the optical axis. Insert
the slider further into the slot or backward to the
desired slider position.
Mechanical Stage (optional)
To install the optional mechanical stage:
1. Install the mechanical according to path
(as shown
in the figure). First, align the edge A of the mechanical
stage with the edge of the flat/plain stage surface.
Align the mechanical stage with the plain stage until
the two set screws in the bottom of the mechanical
stage align with screw holes in the bottom of the plain
stage. Tighten the two set screws.
2. Install the universal holder according to path
(as
shown in the figure). Start by placing the flat universal
holder plate on the plain stage surface. Align the two
screw holes on the universal holder plate with the set
screws on the lateral movement ruler of the mechanical
stage. Tighten the two set screws.
Set screw
Adjustment screws