12
Control Tests
Understanding Out-of-range Control results
WArNINg
The control range applies only to a control result. It only indicates that the test strips and meter are working
properly. Do not use a control result to interpret blood glucose results.
If the control result is out of range, do not use the meter until you solve the problem. Check this list to help
solve the problem.
Troubleshooting Checks
Action
1. Were the test strips or
control solutions
expired?
Discard the test strips or control solution if either is
past the
Use By
date. If the control solution was
opened more than 3 months ago, discard it. Repeat
the control test with an unexpired test strip and an
unexpired control solution.
2. Did you wipe the tip of the control solution
bottle before use?
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
3. Were the caps on the test strip container and
the control solution bottle always closed
tightly?
Replace the test strips or control solution if you think
either was uncapped for some time. Repeat the
control test.
4. Was the test strip used immediately after it
was removed from the test strip container?
Repeat the control test with a new test strip and a
fresh drop of control solution.
5. Were the test strips and control solutions
stored in a cool, dry place?
Repeat the control test with a properly stored test
strip or control solution.
6. Did you follow the directions?
Read Chapter 2, Control Tests, and repeat the control
test.
7. Are you still unsure of the problem?
Contact the ACCU-CHEK Customer Care Service
Center at 1-800-858-8072.
51742_06333630003.indb 12
1/22/13 1:17 PM
Содержание nano 48210
Страница 1: ...Owner s Booklet for Self Testing Only Blood Glucose Meter 51742_06333630003 indb 1 1 22 13 1 17 PM...
Страница 50: ...Notes 48 51742_06333630003 indb 48 1 22 13 1 17 PM...
Страница 51: ...Manual del usuario nicamente para autocontrol Medidor de glucemia 51742_06333630003_ES indb 1 1 25 13 11 32 AM...
Страница 100: ...Notas 48 51742_06333630003_ES indb 48 1 25 13 11 33 AM...