This message/
symbol appears ...
and means:
A test error has occurred.
Turn the meter off and on again. Repeat the test.
There was not enough blood or control solution
on the test strip.
Turn the meter off and on again. Repeat the test
with a larger drop of blood or control solution and
a new test strip.
You forcibly removed the test strip during
testing.
Turn the meter off and on again. Repeat the
test with a new strip, this time ejecting it by
means of the ON/OFF button.
You bent the test strip while applying blood
or control solution or during measurement.
Turn the meter off and on again. Repeat the
test with a new strip.
The optical system cover may not be closed
properly.
Ensure that the optical system cover is firmly
in place.
This message/
symbol appears ...
and means:
The test strip is stuck in the guide.
Turn the meter off. Pull the strip out of the meter.
Open the drum compartment and remove the
drum. Reinsert the drum and close the drum
compartment. The drum symbol and “CTRL”
flash in the display when the meter is turned on.
Perform a quality control check. You can use all
remaining test strips. The 90-day use-by period
will no longer be properly calculated for this
drum. The drum symbol continues to flash in the
display until a new drum is inserted.
It is too warm or too cold for the test, i.e. warmer
than +45 °C or colder than +5 °C. No test result
is displayed or saved.
Turn the meter off. Let the meter and test strips
slowly cool down or warm up. Turn the meter
on and repeat the test. The optimum measuring
temperature is b10 °C and +40 °C.
9.2 Troubleshooting