This message/
symbol appears ...
and means:
If you inserted the second or third drum
from a pack, press the SET button. The
meter uses the bar code from the first
drum. The error message “E-2” disap-
pears. The drum symbol and “CTRL”
flash in the display, requesting you to
perform a quality control check. You can
use all 17 test strips. The drum symbol
continues to flash in the display until a
new drum is inserted.
If you inserted a drum from a single-
drum pack, you can no longer use this
drum. Insert a new drum.
All test strip drums in the same package
have a bar code unique to that pack-
age. Do not try to code the meter using
drums from a different package. If the
test strips and bar code do not belong
together, inaccurate blood glucose re-
sults may be obtained. Check you are
using the correct drums by comparing
their lot codes. The lot code is printed on
the label of the test strip drum next to
the
symbol.
This message/
symbol appears ...
and means:
If the message is still displayed, please
contact Customer Support (for details
see Chapter 19).
CTRL
The 90-day “use-by” period has been exceeded.
The test strips in the drum are unusable.
Insert a new drum.
If you continue to use the drum, carry
out a quality control check. The test
results obtained under these conditions
may only be used for reference purposes
and must not be used for decisions re-
garding therapy. Date and drum symbol
continue to flash in the display until a
new drum is inserted.
The 90-day “use-by” period has been ex-
ceeded. You performed the requested quality
control check and flagged it as such.
When retrieving results: The test strips used
to obtain the stored result were unusable
(more than 90 days old).
The test results obtained under these
conditions may only be used for refer-
ence purposes and must not be used for
decisions regarding therapy.
9.2 Troubleshooting