58
Farbe: PMS 583 – Prüfmittelnummer: 208
59
5
Using the Accu-Chek Integra meter as an
electronic notebook
5.1 Memory
Your Accu-Chek Integra meter has a 100-value memory for saving re-
sults together with time and date (provided the latter are correctly set).
You do not need to save a result specially, the meter does it all for you.
Once all the memory locations are full, the meter automatically over-
writes the oldest entry. All results in memory are consecutively num-
bered from 1 (most recent) to a maximum of 100 (oldest).
In addition to results, your meter also stores all relevant additional
information and display messages. For example, results flagged as
control results (see Chapter 6.2 “Performing a quality control check”)
are always displayed together with “CTRL”. Results obtained at a tem-
perature outside the permitted range are displayed together with a ther-
mometer symbol ( ). Please refer to Chapter 9 “Display messages and
troubleshooting” for a complete list of display messages and symbols.
5.2 Viewing test results
With the meter turned off, press the MEMO button.
The meter turns on without advancing a test strip
from the drum. The most recent test result and
“MEMO” are displayed.
i
Only
turn on the meter with the MEMO button when you would
like to review results held in memory.
If you turn on the meter with the ON/OFF button and there is a
drum in the meter, a test strip will be advanced.
Press the SET button to move towards the
“oldest” result stored.
Press the MEMO button to scroll forward to the
“most recent” result again.
You can hold down the buttons to scroll through your results more
quickly.
Press the ON/OFF button to turn the meter off.
A test strip is not ejected.
i
If you did not set the time and date, your
results will be displayed with the number
of their memory location.
MEMO
mmol/L
MEMO
mmol/L
MEMO
mmol/L