background image

HP 300-121 

Page 4 of 5 

 

Reviewed/Revised 06/26/13 

15.  After the sample has been obtained, apply gentle pressure to the puncture with a clean gauze square 

or cotton ball site for several minutes. 

16.  The measurement is complete when the result is displayed on the screen.  Depending upon how high 

or low the result is, it may appear in a numeric or non-numeric format.  See 

Reporting Results

 section 

below for interpretation of each result format. 

17.  Remove the test strip and dispose of it in plain trash.

 

18.  Touch 

 

to  enter  up  to  three  appropriate  comment(s)  as  required  in  the  “General  Policies” 

section of this manual.

 

19.  Touch the 

 button to confirm the result and send the result from the meter wirelessly or place the 

meter  in  the  base  unit  to  send  the  result  and  record  the  result  into  the  electronic  data  management 
system. The base unit also charges the meter. 

20. 

Documentation of the result in the patient’s electronic medical record occurs automatically, as long as 
the patient’s FIN number is entered correctly. 

21.  Follow up on any results that exceed critical or reportable limits according to policy. 

22.  Clean and disinfect as necessary with Clorox wipes. 

23.  Error messages: 

 

If the error message “

Strip Defect Error

” appears on the display, the test strip may be defective or 

the blood glucose result may be extremely low and below the meter’s measurement range.  Refer 
to the test strip package insert, perform a quality control test using a new test strip, review  proper 
testing procedure, and repeat the blood glucose test, or follow y

our facility’s testing policy. 

 

If the meter displays “

Type Bad Dose,

” there may be insufficient amount of blood on the test strip. 

Repeat the test using a new test strip, ensuring proper sample application, or refer to the test strip 
package insert. 

 

Guidance for Interpreting On-Screen Message and Error Codes: 

All  error  messages  displayed  by  the  system  have  a  letter  identifying  the  message  type,  a 
number  and  a  description  of  the  error  to  help  the  operator  take  action  to  resolve  the  problem. 
The different message types are in the table below.  

Identifies  the  notification  as  an  Error.    The  information  notifies  the 
operator that an error has occurred. 

Identifies the notification as a Warning.  The information does not block 
the  operator  from  continuing,  but  rather  gives  the  operator  information 
that may suggest an alternate workflow is required. 

Identifies  the  notification  as  Informational  only.    Informational 
notifications  present the  operator  with contextual  information, and allow 
the operator to proceed after confirming the notification. 

Identifies  a  Decision  point.    Decision  notifications  provide  the  operator 
with a choice based on contextual information. 

Linearity Testing Procedure 

Linearity  testing,  also  known  as  calibration  verification,  is  a  means  of  testing  the  ACCU-CHEK  Inform  II 
system’s ability to give correct readings through a broad range of glucose levels.  It is used as a periodic 
check  on  system  performance.    Linearity  testing  is  performed  on  10%  of  meters  during  new  instrument 
implementation, then on all subsequent meters or as part of troubleshooting related to questionable meter 
performance. 

Proficiency Testing Procedure 

Participation  in  proficiency  testing  is  part  of  the  quality  assurance  program  to  ensure  proper  on-going 
performance of the ACCU-CHEK Inform II system and the competence of staff performing tests.  At least 
10%  of  meters  are  tested  during  each  proficiency  cycle  by  staff  routinely  using  the  device.    Results  are 
submitted to CAP without consulting with any other survey subscribers.  The survey report is reviewed by 

Содержание Inform II Competency

Страница 1: ...NORTHERN ARIZONA HEALTHCARE ACCU CHEK Inform II Competency 2013 Sponsored by the Education Department...

Страница 2: ...your clinicals 2 Complete the Post Test and use the Answer Sheet provided to record your responses to the Post Test Print out the answer sheet and sign the statement showing that you have read the Gui...

Страница 3: ...erstick and neonate heelstick whole blood b The following anticoagulants are acceptable do not use any other anticoagulants for meter testing i Lithium or Sodium Heparin EDTA c Sufficient sample size...

Страница 4: ...oratory DO NOT attempt to perform tests under another operator s ID c From the Main Menu touch Control Test d Select the level of control that you wish to test e Confirm that the meter is coded calibr...

Страница 5: ...Patient Testing a Wash hands and don personal protective equipment gloves gowns etc as required by infection control and isolation policies and procedures b Assemble the finger stick equipment the me...

Страница 6: ...ent is complete when the result is displayed on the screen Depending upon how high or low the result is it may appear in a numeric or non numeric format See Interpretation of Results section below for...

Страница 7: ...ent result obtained by the ACCU CHEK Inform ll does not appear to fit the clinical picture repeat the QC and patient specimen A You must use the comment Repeat will not post on the repeat test The pat...

Страница 8: ...nd results as soon as a wireless connection is established The meter will connect to Cerner every 10 minutes and transfer new data when a wireless connection is available even if the meter is powered...

Страница 9: ...elson MT ASCP 7 12 Karen McMullin Director 7 12 4 13 Approved by Title Date Susanne Maiden CNO VP 6 13 Committee Approval Date Policy Procedure 10 06 5 18 07 Nursing Practice 8 25 10 GOP 9 21 10 7 17...

Страница 10: ...anticoagulants for meter testing a Lithium or Sodium Heparin EDTA 3 Refer to Mosby s Nursing Skills for detailed instructions on how to perform a finger or heel stick MATERIALS Accu Chek Inform II met...

Страница 11: ...ated to the same test strip code that is printed on the test strip vial by scanning the test strip vial s barcode Contact the laboratory if unable to confirm the correct test strip code 6 The meter wi...

Страница 12: ...ID by means of scanning the barcode on your identification badge NOTE If the operator ID you enter is not accepted attempt to re enter it If it is still rejected contact your supervisor or Point of Ca...

Страница 13: ...ensuring proper sample application or refer to the test strip package insert Guidance for Interpreting On Screen Message and Error Codes All error messages displayed by the system have a letter ident...

Страница 14: ...0 mg dL and for newborns 30 mg dL and 300 mg dL Patient test results that exceed critical limits must be called to the patient s provider and then documented in the patient electronic medical records...

Страница 15: ...If it is not the correct amount it will give an error code a True b False 3 After applying the sample to the test strip and the measurement begins an error message occur this means a The sample is in...

Страница 16: ...e more times to be sure d None of the above 9 What is true of test strip handling a Test strips do not have an expiration date b Remove the test strip from the vial and immediately recap the vial c Un...

Страница 17: ...rop of blood to the front edge of the Accu chek Inform ll test strip a True b False 14 Only capillary whole blood samples should be used when testing the Accu chek Inform ll meter and the Accu chek In...

Страница 18: ...d 7 a b c d 8 a b c d 9 a b c d 10 a b 11 a b c d e 12 a b c d e 13 a b 14 a b Instructor s Name ___________________________________________________ The undersigned certifies as follows I have read t...

Страница 19: ...onfirmation 30 or 500 mg dL 6 months to Adult 30 or 500mg dL 0 6months States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning State...

Страница 20: ...ral laboratory for confirmation 30 or 500 mg States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning States proper infection control...

Отзывы: