Page 2 of 7   200.02 Point of Care ACCU-CHEK Inform ll Blood Glucose Monitoring System 

 

m.  Do not get liquid into the strip port!  If liquid does get into the test strip port, immediately 

dry the components with a dry cloth or gauze pad. If solution is allowed to collect in any 
meter opening, severe damage to the system can occur.  If you suspect that moisture 
may have entered the strip, perform glucose control testing. 

n.  Do not wipe the electrical connectors on the back of the base unit. 
o.  Do not use any cleaning and disinfecting product other than that which is recommended 

by the manufacturer, identified in this procedure and provided through normal 
procurement policies and procedures. 

 

f. 

Quality Control Testing 

a.  Two  levels  of  quality  control  material  (Low  and  High)  with  known  values  must  be 

performed every use day.   

b.  If a facility is closed quality control will be done when the facility opens, prior to testing of 

the first patient.   

c.  Results  of  both  levels  of  quality  control  must  be  within  the  acceptable  ranges  before 

patient testing is done. 

d.  Failure to successfully run quality control within the 24 hour window will result in the  

   ACCU-CHEK Inform ll locking out all patient testing. 

e.  Quality control must also be performed in the following circumstances: 

i.  When a new vial of test strips is opened. 

ii.  If the cap has been left off a vial of test strips. 

iii.  If the meter has been dropped. 

iv.  When questioning patient results or the performance of the meter. 

 

g.  Procedure for Quality Control Testing 

a.  Turn the meter on by pressing the ON/OFF button located below the center of the touch 

screen. 

b.  Enter the operator ID by means of scanning the barcode on the identification band.   

NOTE:

 If the operator ID you enter is not accepted, attempt to re-enter it.  If it is still 

rejected, contact the laboratory. 

 DO NOT 

attempt to perform tests under another 

operator’s ID.  

c.  From the 

Main Menu

, touch

 Control Test

.

 

d.  Select the level of control that you wish to test.

 

e.  Confirm that the meter is coded (calibrated) to the same test strip code that is printed on 

the test strip vial by scanning the test strip vial’s barcode. Contact the laboratory if unable 
to confirm the correct test strip code.

 

f. 

The meter will display a picture of a test strip with a downward flashing arrow on the 
meter indicating that you are ready to insert a test strip into the meter. Remove a test 
strip from the vial and immediately recap the vial. Insert the test strip into the meter in the 
direction of the arrows and with the “ACCU-CHEK” lettering facing upward. The meter will 
display a flashing drop above the test strip icon when the test strip is properly inserted 
indicating that you are ready to apply control solution.  

 

g.  Apply control solution to the front edge of the test strip. The solution will fill the yellow 

sample chamber by capillary action. Do not apply sample to the top of the test strip. Once 
sufficient sample has been detected, the measurement begins. An hourglass icon 
indicates that the measurement is in progress. You will get an error message if the 
sample is insufficient. If this occurs, you will need to repeat the test.

 

h.  The measurement is complete when the result is displayed on the meter screen.  See the 

“Interpretation of Results” and “Troubleshooting” sections below for guidance on what to 
do if your result is “Fail” or shows an “out of range” message associated with the result. 

i. 

Remove the test strip and dispose of it in plain trash.

 

j. 

Touch the comment button ( 

) to enter an appropriate comment(s) as required in 

the “General Policies” section of this manual.

 

k.  Touch the 

  button to confirm the result and send the result from the meter 

wirelessly or place the meter in the base unit to send the result and record the result into 
the electronic data management system. The base unit also charges the meter. 

 
 
 

Содержание Inform II Competency

Страница 1: ...NORTHERN ARIZONA HEALTHCARE ACCU CHEK Inform II Competency 2013 Sponsored by the Education Department...

Страница 2: ...your clinicals 2 Complete the Post Test and use the Answer Sheet provided to record your responses to the Post Test Print out the answer sheet and sign the statement showing that you have read the Gui...

Страница 3: ...erstick and neonate heelstick whole blood b The following anticoagulants are acceptable do not use any other anticoagulants for meter testing i Lithium or Sodium Heparin EDTA c Sufficient sample size...

Страница 4: ...oratory DO NOT attempt to perform tests under another operator s ID c From the Main Menu touch Control Test d Select the level of control that you wish to test e Confirm that the meter is coded calibr...

Страница 5: ...Patient Testing a Wash hands and don personal protective equipment gloves gowns etc as required by infection control and isolation policies and procedures b Assemble the finger stick equipment the me...

Страница 6: ...ent is complete when the result is displayed on the screen Depending upon how high or low the result is it may appear in a numeric or non numeric format See Interpretation of Results section below for...

Страница 7: ...ent result obtained by the ACCU CHEK Inform ll does not appear to fit the clinical picture repeat the QC and patient specimen A You must use the comment Repeat will not post on the repeat test The pat...

Страница 8: ...nd results as soon as a wireless connection is established The meter will connect to Cerner every 10 minutes and transfer new data when a wireless connection is available even if the meter is powered...

Страница 9: ...elson MT ASCP 7 12 Karen McMullin Director 7 12 4 13 Approved by Title Date Susanne Maiden CNO VP 6 13 Committee Approval Date Policy Procedure 10 06 5 18 07 Nursing Practice 8 25 10 GOP 9 21 10 7 17...

Страница 10: ...anticoagulants for meter testing a Lithium or Sodium Heparin EDTA 3 Refer to Mosby s Nursing Skills for detailed instructions on how to perform a finger or heel stick MATERIALS Accu Chek Inform II met...

Страница 11: ...ated to the same test strip code that is printed on the test strip vial by scanning the test strip vial s barcode Contact the laboratory if unable to confirm the correct test strip code 6 The meter wi...

Страница 12: ...ID by means of scanning the barcode on your identification badge NOTE If the operator ID you enter is not accepted attempt to re enter it If it is still rejected contact your supervisor or Point of Ca...

Страница 13: ...ensuring proper sample application or refer to the test strip package insert Guidance for Interpreting On Screen Message and Error Codes All error messages displayed by the system have a letter ident...

Страница 14: ...0 mg dL and for newborns 30 mg dL and 300 mg dL Patient test results that exceed critical limits must be called to the patient s provider and then documented in the patient electronic medical records...

Страница 15: ...If it is not the correct amount it will give an error code a True b False 3 After applying the sample to the test strip and the measurement begins an error message occur this means a The sample is in...

Страница 16: ...e more times to be sure d None of the above 9 What is true of test strip handling a Test strips do not have an expiration date b Remove the test strip from the vial and immediately recap the vial c Un...

Страница 17: ...rop of blood to the front edge of the Accu chek Inform ll test strip a True b False 14 Only capillary whole blood samples should be used when testing the Accu chek Inform ll meter and the Accu chek In...

Страница 18: ...d 7 a b c d 8 a b c d 9 a b c d 10 a b 11 a b c d e 12 a b c d e 13 a b 14 a b Instructor s Name ___________________________________________________ The undersigned certifies as follows I have read t...

Страница 19: ...onfirmation 30 or 500 mg dL 6 months to Adult 30 or 500mg dL 0 6months States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning State...

Страница 20: ...ral laboratory for confirmation 30 or 500 mg States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning States proper infection control...

Отзывы: