HP 300-121 

Page 5 of 5 

 

Reviewed/Revised 06/26/13 

the Laboratory Director/Medical Director or Designee

 

Appropriate corrective action is taken in the event of 

survey failures. 

 

REPORTING RESULTS 

Data is transferred from the meter into Cerner as follows:  

1. 

Wireless  Data  Transfer:

    Results  will  automatically  transfer  wirelessly  into  Cerner  if  a  wireless 

connection  is  currently  available  upon  completion  of  a  test.    A  wireless  connection  is  activated  when 

the testing operator touches 

 upon completing a test.  The meter will connect to Cerner every 10 

minutes and transfer new  data  when  a  wireless connection is available even  if the meter is  powered 
off.  This connection gives the meter the most up to date settings and ADT information when available. 

2. 

Wired Data Transfer:

  Results are transferred into Cerner by placing the meter into a base unit that is 

connected  to  the  hospital  network  through  an  RJ45  connection  connected  to  the  hospital  network).  
When properly docked the meter will automatically transfer test results.  The following series of screen 
displays will appear to confirm that data  transfer is occurring: 

Critical limits

 for adult patient tests are 

≤45 mg/dL and ≥450 mg/dL

 and for newborns

, ≤30 mg/dL and 

≥300  mg/dL

.   

Patient  test  results  that  exceed  critical  limits  must  be  called  to  the  patient’s  provider  and 

then documented in the patient electronic medical records. 

Reportable limits

 for patient testing are 

≤30 mg/dL and ≥500 mg/dL

.  Results that exceed the reportable 

limits of the system will be displayed as RR HI/RR LO or HI/LO, the measurement range default.  Patient 
test results that exceed reportable limits must have follow-up testing performed in the laboratory. 
 
ATTACHMENTS 

N/A 

 
REFERENCES 

 

ACCU-

CHEK 

Inform 

II 

User’s 

Guide. 

 

Retrieved 

from: 

www.poc.roche.com

  and 

www.accu-checkinformii.com

  websites,  ©  2012-2013, 

Roche Diagnostics.  All rights reserved. 

Mosby’s Nursing Skills, 2013. 

 

Blood Glucose Monitoring.  Retrieved from

 

http://mns.elsevierperformancemanager.com/NursingSkills/ContentPlayer/Skill
ContentPlayerIFrame.aspx?KeyId=430&Id=GN_43_12&Section=1&bcp=Sear
chOp~0~glucose~False&IsConnect=False

 

 

Blood Glucose Monitoring (Pediatric).  Retrieved from

 

http://mns.elsevierperformancemanager.com/NursingSkills/ContentPlayer/Skill
ContentPlayerIFrame.aspx?KeyId=855&Id=CCP_133&Section=1&bcp=Searc
hOp~0~glucose~False&IsConnect=False

 

 

Heel Stick (Neonatal).  Retrieved from 

http://mns.elsevierperformancemanager.com/NursingSkills/ContentPlayer/Skill
ContentPlayerIFrame.aspx?KeyId=1233&Id=NN_003&Section=1&bcp=Searc
hOp~0~heelstick~False&IsConnect=False

 

 

Содержание Inform II Competency

Страница 1: ...NORTHERN ARIZONA HEALTHCARE ACCU CHEK Inform II Competency 2013 Sponsored by the Education Department...

Страница 2: ...your clinicals 2 Complete the Post Test and use the Answer Sheet provided to record your responses to the Post Test Print out the answer sheet and sign the statement showing that you have read the Gui...

Страница 3: ...erstick and neonate heelstick whole blood b The following anticoagulants are acceptable do not use any other anticoagulants for meter testing i Lithium or Sodium Heparin EDTA c Sufficient sample size...

Страница 4: ...oratory DO NOT attempt to perform tests under another operator s ID c From the Main Menu touch Control Test d Select the level of control that you wish to test e Confirm that the meter is coded calibr...

Страница 5: ...Patient Testing a Wash hands and don personal protective equipment gloves gowns etc as required by infection control and isolation policies and procedures b Assemble the finger stick equipment the me...

Страница 6: ...ent is complete when the result is displayed on the screen Depending upon how high or low the result is it may appear in a numeric or non numeric format See Interpretation of Results section below for...

Страница 7: ...ent result obtained by the ACCU CHEK Inform ll does not appear to fit the clinical picture repeat the QC and patient specimen A You must use the comment Repeat will not post on the repeat test The pat...

Страница 8: ...nd results as soon as a wireless connection is established The meter will connect to Cerner every 10 minutes and transfer new data when a wireless connection is available even if the meter is powered...

Страница 9: ...elson MT ASCP 7 12 Karen McMullin Director 7 12 4 13 Approved by Title Date Susanne Maiden CNO VP 6 13 Committee Approval Date Policy Procedure 10 06 5 18 07 Nursing Practice 8 25 10 GOP 9 21 10 7 17...

Страница 10: ...anticoagulants for meter testing a Lithium or Sodium Heparin EDTA 3 Refer to Mosby s Nursing Skills for detailed instructions on how to perform a finger or heel stick MATERIALS Accu Chek Inform II met...

Страница 11: ...ated to the same test strip code that is printed on the test strip vial by scanning the test strip vial s barcode Contact the laboratory if unable to confirm the correct test strip code 6 The meter wi...

Страница 12: ...ID by means of scanning the barcode on your identification badge NOTE If the operator ID you enter is not accepted attempt to re enter it If it is still rejected contact your supervisor or Point of Ca...

Страница 13: ...ensuring proper sample application or refer to the test strip package insert Guidance for Interpreting On Screen Message and Error Codes All error messages displayed by the system have a letter ident...

Страница 14: ...0 mg dL and for newborns 30 mg dL and 300 mg dL Patient test results that exceed critical limits must be called to the patient s provider and then documented in the patient electronic medical records...

Страница 15: ...If it is not the correct amount it will give an error code a True b False 3 After applying the sample to the test strip and the measurement begins an error message occur this means a The sample is in...

Страница 16: ...e more times to be sure d None of the above 9 What is true of test strip handling a Test strips do not have an expiration date b Remove the test strip from the vial and immediately recap the vial c Un...

Страница 17: ...rop of blood to the front edge of the Accu chek Inform ll test strip a True b False 14 Only capillary whole blood samples should be used when testing the Accu chek Inform ll meter and the Accu chek In...

Страница 18: ...d 7 a b c d 8 a b c d 9 a b c d 10 a b 11 a b c d e 12 a b c d e 13 a b 14 a b Instructor s Name ___________________________________________________ The undersigned certifies as follows I have read t...

Страница 19: ...onfirmation 30 or 500 mg dL 6 months to Adult 30 or 500mg dL 0 6months States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning State...

Страница 20: ...ral laboratory for confirmation 30 or 500 mg States emergency ID when patient is unidentified 0911911 Verbalizes how to enter a custom comment patient s name 5 Cleaning States proper infection control...

Отзывы: