Other Information
DISPLAY
ACTION
CAUSE
The meter shuts off
after three seconds.
Clock has stopped.
Clock is slow after
the meter is
turned on.
Insufficient battery power – no
further tests can be performed.
The meter was exposed to tem-
peratures below +14° F (-10° C)
and the batteries are starting to
freeze.
Temperature was outside the
indicated range during testing.
It is too warm or too cold for the
test, i.e. warmer than +104° F
(+40° C) or colder than +50° F
(+10° C).
Insert two new batteries (type AAA).
Always store the meter above +14° F (-10° C).
The displayed result may only be used for reference pur-
poses. It is saved along with the thermometer symbol.
To perform an accurate test, let the meter and test drum
slowly cool down or warm up and repeat the test. The
optimum measuring temperature is b50° F
(+10° C) and +104° F (+40° C).
No test result is displayed or saved. Let the meter and
test drum slowly cool down or warm up and repeat the
test. The optimum measuring temperature is between
+50° F (+10° C) and +104° F (+40° C).
▲
EXAMPLE
361-11264 CMPCT Manual (Eng) 8/8/02 3:46 PM Page 96
Содержание Compact
Страница 1: ......
Страница 18: ...31 Getting Started ...
Страница 40: ...75 Other Information You can hold down the buttons to scroll through your results more quickly EXAMPLE ...
Страница 41: ...77 Other Information After the MEM button is pressed the meter then turns on without advancing a test strip ...
Страница 44: ...83 Other Information Make sure the software is ready to accept data from the meter ...
Страница 48: ...91 Other Information The cover must be snapped down completely or the strip won t advance properly ...
Страница 59: ...INSERT START Spanish Cover HERE Signature pg 115 ...