Symbols, error messages and troubleshooting
12
Error message
Meaning
The test strip that was advanced is
defective.
Hold the meter upright so the test strip
is pointing downwards. Press the Test
button to turn the meter off and allow
the test strip to drop out. Press the
Test button to turn the meter on again
and repeat the test.
You pulled the test strip out of the
meter.
Press the Test button to turn the meter
off. Only remove the test strip from the
meter by pressing the Test button to
turn the meter off.
There is condensation in the meter.
Press the Test button to turn the meter
off. Leave it to dry at room
temperature.