191
Symbols, error messages and troubleshooting
12
Error message
Meaning
Turn the meter off and then on again
by pressing the Test button. Repeat
the test with a new test strip.
You drew up blood or control solution
too soon with the test strip, i.e. before
the test strip symbol and drop symbol
were flashing in the display.
Turn the meter off and then on again
by pressing the Test button. Repeat
the test with a new test strip. Only
draw up blood or control solution with
the test strip when the test strip
symbol and drop symbol are flashing
in the display.
You pulled the test strip out of the meter
during testing.
Turn the meter off and then on again by
pressing the Test button. Repeat the test
with a new test strip. Only remove the test
strip from the meter by pressing the Test
button to turn the meter off.
man_06919570001_01_EnCan.indd 191