Symbols, display messages and troubleshooting
157
Symbols, display messages and troubleshooting
10
This message/symbol
appears…
and means:
You pulled the test strip out of the
meter during testing.
Press the Test button to turn the
meter off and then on again. Repeat
the test with a new test strip. Always
remove the test strip by pressing the
Test button to turn the meter off.
This message/symbol
appears…
and means:
The test strip is stuck.
The test strip drum does not rotate.
Press the Test button to turn the
meter off. Open the drum compart-
ment and, if necessary, pull out
the test strip. Press the red drum
release button and remove the test
strip drum. Insert the test strip drum
and close the drum compartment.
The meter then rotates the drum
and reads the bar code. The drum
symbol and bottle symbol then flash
in the display. The 90-day use-by
period will no longer be properly
calculated for this drum. Perform a
quality control check. You may use
up all the test strips remaining in the
drum. The drum symbol continues
to flash in the display until you have
inserted a new drum.