24
Farbe: P 249 – Prüfmittelnummer: 126 X
Initial steps before testing
25
Initial steps before testing
2
2.4 Performing a full display check
To check that all of the elements are properly displayed, you should
carry out a full display test.
Press and hold down the M button.
The meter turns on. As long as you hold down the M button, all the dis-
play elements are visible.
Compare the display screen of your meter with the one shown
below.
mmol/L meter
mg/dL meter
or
All elements displayed
If any of the display elements are missing or the unit shown for the
blood glucose result is not the correct one, please ask your dealer to
exchange the meter.
Release the M button.
The standard display test screen that is shown each time the meter is
turned on appears briefly. When it disappears, the meter is in memory
mode. As no results have yet been saved, three bars are displayed.
Press the Test button to turn the meter off.
i
The meter automatically turns off after 60 seconds even if you
keep the M button pressed.
i
You can perform a full display test whenever you wish. If there
are already results saved in memory, the most recently saved
result is displayed when you release the button.