68
8.
When the test strip has enough
control solution, the Analyzing
screen appears.
j
Wipe the tip of the bottle with a
tissue and then cap the bottle
tightly.
Analyzing
bG Test
9.
10:02 am
2 Feb 08
Control Result
Menu
Level
mg/dL
50
The control result appears on
the display.
j
Select Level to display the
control solution level pop-up
menu.
7.
j
Squeeze the bottle until a tiny
drop forms at the tip.
j
Touch the drop to the
front
edge
of the yellow window
of the test strip. Do not put
control solution on top of the
test strip.
Apply Sample
bG Test
45702_std.indb 68
10/29/09 1:42:12 PM