8
Display Shows Possible Cause(s)
Possible Solution(s)
carbohydrate Ratio
j
carbohydrate ratio is outside
of the acceptable meter
range.
j
check your entries and contact your healthcare
professional to determine the appropriate settings.
code key Missing
j
The meter is not coded or the
code key is not inserted.
j
Turn off the meter and recode it. See chapter 2,
“Testing Your Blood Sugar.”
Diary entry used
For Advice
j
The diary entry selected has
been used for bolus advice
and modifications are not
allowed.
j
Carefully confirm all information involving bolus
advice. You cannot modify bolus advice-related
entries on the meter.
Diary Results
expired
j
Bolus advice is no longer
available for this result.
j
Review data that was saved in My Data. Then,
decide to modify data, add data, or begin a new
bolus advice session.