Accora Configura Advance SC3 Скачать руководство пользователя страница 8

 

8

 

 

14.

 

Locate the two sidepiece fixing bars into the chair 
frame.  Push the sidepiece all the way in (Fig 10) 
and release the plunger by rotating it until the cap 
engages with the base (Fig 11).

 

 

 

 

 

 

15.

 

Ensure the plunger is engaged by moving the 
sidepiece in and out 

 the sidepiece should not 

move. 

16.

 

Tighten the hand screw to secure the sidepiece. 

17.

 

Repeat steps 12 to 16 to fit the other sidepiece. 

18.

 

Carry out a functional check (see section 11). 

 
 
 
 
 
 
 
 
 
 
 

7.

 

Chair adjustment 

 

The Configura Advance chair can be adjusted in the 
following areas: 

1.

 

Chair Width range

 

 The chair width has 4 widths 

that can be set, adjusted using the armrest locking 
plungers and hand screws.  The available width 
dimensions are: 

400 mm (16”)

 

450 mm 

(18”)

 

500 mm (20”)

 

550 mm (22”)

 

2.

 

Chair Depth range

 

 The chair depth can be 

adjusted in 9 increments from approximately 422 

mm (16”) to 5

12 mm (20

”). The depth is adjusted 

using a lever and slide mechanism. 

3.

 

Armrest Height range

 

 The armrest height can be 

set at 

165mm (6.5”), 200mm (8”) 

or 

225mm (9”). 

The height is adjusted by changing the armrest 
bracket fitting position. 

 

4.

 

Footplate Height range 

 

The

 

footplate height can 

be adjusted from 330mm 

(13”) to 533mm (21”).  

The height is adjusted using two plungers. 

5.

 

Chair Seat Height range 

 

The

 

Advance chair 

frame can be adjusted to give seat heights from 
the floor 

of 615mm (24.2”) and 540mm (21.2”).

 

Chair Width adjustment 

1.

 

Make sure a patient is not in the chair. 

2.

 

Make sure the brakes on all 3 locking castors are 
locked on (See section 10).   

3.

 

On one side of the chair, underneath the base of 
the sidepiece, locate the locking plunger (Fig 8, 
Item 1). 

4.

 

Pull the plunger out and rotate 90 degrees to lock 
the plunger open (Fig 9). 

5.

 

Loosen the hand screw (Fig 8, Item 2). 

6.

 

With two hands, pull or push the sidepiece until the 
desired measurement is achieved. The 
measurement can be read just inside the 
sidepiece at the front between the seat and the 
sidepiece (See Fig 12). 

7.

 

Pull the plunger out, rotate 90 degrees and let go 
to lock the plunger in position (Fig 11). 

Figure 9

 

Figure 10 

Figure 11

 

WARNING 

 
Adjustment MUST be carried out by suitably 
trained and qualified personnel. 
A clinical assessment and risk assessment MUST 
be carried out by suitably trained and 
qualified personnel before adjustments are 
made to the chair. 
Do not adjust the width, depth or height of 
the chair when a patient or user is in the chair. 

 

 

 

 

Содержание Configura Advance SC3

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC3 001US Rev 02 March 2021...

Страница 2: ...fied as a Class 1 Medical Device in accordance with the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation to this medical device affecting the user or the pati...

Страница 3: ...pump cable could lead to dangerous electrical hazards if squeezed or crushed between moving parts The chair must not be used if there is any visible damage to any cables 12 Do not move the chair when...

Страница 4: ...tended for use in primary healthcare environments where supervision can be provided if necessary for example long term residential care nursing homes or convalescent units The Configura Advance chair...

Страница 5: ...1 050 Dropdown sidepieces and covers set of 2 DRPARM 0 SC1 250 Dropdown armrests 400 550mm 16 22 set of 2 DRPARM 0 SC1 050 Dropdown armrests 350 500mm 14 20 set of 2 DRPARM 0 SC1 150 Dropdown armrests...

Страница 6: ...with the clevis at the back of the main frame Fig 3 Item 1 Ensure the Tilt in Space operating cable Fig 3 Item 4 is routed between the backrest and the backrest actuator assembly 6 Insert the clevis p...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...te Height range The footplate height can be adjusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers 5 Chair Seat Height range The Advance chair frame can be adjusted to give seat...

Страница 9: ...h the other hand The depth setting can be read off the scale on the back part of the seat frame Fig 14 Item 1 Figure 13 5 Release the depth adjustment lever and move the backrest in and out slightly t...

Страница 10: ...footplate See section 8 and place the chair carefully on its front 4 Using a 13mm spanner and 5mm Allen key undo and remove the Nylock nut and shoulder bolt Fig 19 Item 1 and Item 2 Keep the bolt was...

Страница 11: ...ig 24 Items 1 2 into the two sockets either side of the footplate height adjustment frame Fig 21 Items 3 4 Push the footplate into the tubes until the plungers latch with an audible click 3 Ensure the...

Страница 12: ...can be used to restrict the use of the backrest angle adjustment 1 Tilt in Space The tilt in space position can adjust the seat angle from 7 degrees to 37 degrees from the horizontal position see pag...

Страница 13: ...e 7 If locking out the backrest is required return the locking lever to the locked position If the locking lever is too loose or too tight it may be necessary to rotate the whole lever clockwise or an...

Страница 14: ...to press Care must be taken to ensure the castor brakes are always locked when the chair is in use being assembled or dismantled so that the chair does not move accidentally Individually braked 125mm...

Страница 15: ...r can be braked at the same time using the foot lever 1 To lock the central brake press the rear red part of the foot lever down until the brake locks Figure 31 shows the brake locked 2 To unlock the...

Страница 16: ...s profiled headrest external laterals 3 Clean as described in the Cleaning Information 13 Troubleshooting Problem Possible solution s Backrest will not recline manual version Ensure the backrest locko...

Страница 17: ...tplate securely attached 12 Are the plungers securing the sidepieces locked and secure 13 Is there any risk of entrapment or patient injury 14 Are any accessory electrical cables pinched crushed or da...

Страница 18: ...t IPX4 Degree of protection against liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in acco...

Страница 19: ...19 This page intentionally left blank...

Страница 20: ...20 IFU SC3 001US REV 02...

Отзывы: