15.

 

Daily inspection 

 

Daily visual inspection is strongly recommended and 
may be carried out by carer, user or another person.  
 

The following checks must be carried out: 
 
1.  Does the chair operate as per its intended purpose 

without unexpected noise or motion? 

2.  Are there any signs of abuse or excessive wear? 
3.  Is any fabric torn or damaged? Pay particular 

attention to the seat overlay. 

4.   Check that the overlay is secured to the frame with 

the press studs. 

5.  Are all fixtures and fittings tight and secure? 
6.  Does the chair frame appear stable and secure? 
7.  Are all accessories fitted in line with the accessory 

manufacturer or accessory supplier’s instructions?

 

8.  Are all the castor brakes in the locked position? 
9.  Are all electrical cables for accessories, (e.g. air 

pump) secured and routed to prevent damage? 

10. Is the area around, above and below chair clear of 

possible obstruction? 

11. Is the footplate securely attached? 
12. Are the plungers securing the sidepieces locked 

and secure? 

13. Is there any risk of entrapment or patient injury? 
14. Are any accessory electrical cables pinched, 

crushed or damaged in any way? 

 
If any damage, performance issue or cause for 
concern is noted during this inspection the chair should 
be withdrawn from service and appropriate steps 
should be taken. 
 

16.

 

General maintenance

 

 

 

 

Repairs to the Configura Advance chair must be 
carried by suitably trained and qualified personnel. 
 
After any maintenance has taken place a functionality 
check must be carried out (see section 11). 
 
 

17.

 

Guarantee 

 

Model  

Configura Advance chair 

Warranty period 

2 years 

 
 

18.

 

Disposal 

 

In the event of the disposal of materials from the chair, 
end-of-life parts must be disposed of in accordance 
with current environmental regulations

WARNING 

 

Maintenance MUST be carried out by suitably 
trained and qualified personnel. 
All functions MUST be tested and approved 
after maintenance by suitably trained and 
qualified personnel.  
Do not carry out maintenance with service 
user or patient on the chair.  
 

Содержание Configura Advance SC3

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC3 001US Rev 02 March 2021...

Страница 2: ...fied as a Class 1 Medical Device in accordance with the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation to this medical device affecting the user or the pati...

Страница 3: ...pump cable could lead to dangerous electrical hazards if squeezed or crushed between moving parts The chair must not be used if there is any visible damage to any cables 12 Do not move the chair when...

Страница 4: ...tended for use in primary healthcare environments where supervision can be provided if necessary for example long term residential care nursing homes or convalescent units The Configura Advance chair...

Страница 5: ...1 050 Dropdown sidepieces and covers set of 2 DRPARM 0 SC1 250 Dropdown armrests 400 550mm 16 22 set of 2 DRPARM 0 SC1 050 Dropdown armrests 350 500mm 14 20 set of 2 DRPARM 0 SC1 150 Dropdown armrests...

Страница 6: ...with the clevis at the back of the main frame Fig 3 Item 1 Ensure the Tilt in Space operating cable Fig 3 Item 4 is routed between the backrest and the backrest actuator assembly 6 Insert the clevis p...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...te Height range The footplate height can be adjusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers 5 Chair Seat Height range The Advance chair frame can be adjusted to give seat...

Страница 9: ...h the other hand The depth setting can be read off the scale on the back part of the seat frame Fig 14 Item 1 Figure 13 5 Release the depth adjustment lever and move the backrest in and out slightly t...

Страница 10: ...footplate See section 8 and place the chair carefully on its front 4 Using a 13mm spanner and 5mm Allen key undo and remove the Nylock nut and shoulder bolt Fig 19 Item 1 and Item 2 Keep the bolt was...

Страница 11: ...ig 24 Items 1 2 into the two sockets either side of the footplate height adjustment frame Fig 21 Items 3 4 Push the footplate into the tubes until the plungers latch with an audible click 3 Ensure the...

Страница 12: ...can be used to restrict the use of the backrest angle adjustment 1 Tilt in Space The tilt in space position can adjust the seat angle from 7 degrees to 37 degrees from the horizontal position see pag...

Страница 13: ...e 7 If locking out the backrest is required return the locking lever to the locked position If the locking lever is too loose or too tight it may be necessary to rotate the whole lever clockwise or an...

Страница 14: ...to press Care must be taken to ensure the castor brakes are always locked when the chair is in use being assembled or dismantled so that the chair does not move accidentally Individually braked 125mm...

Страница 15: ...r can be braked at the same time using the foot lever 1 To lock the central brake press the rear red part of the foot lever down until the brake locks Figure 31 shows the brake locked 2 To unlock the...

Страница 16: ...s profiled headrest external laterals 3 Clean as described in the Cleaning Information 13 Troubleshooting Problem Possible solution s Backrest will not recline manual version Ensure the backrest locko...

Страница 17: ...tplate securely attached 12 Are the plungers securing the sidepieces locked and secure 13 Is there any risk of entrapment or patient injury 14 Are any accessory electrical cables pinched crushed or da...

Страница 18: ...t IPX4 Degree of protection against liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in acco...

Страница 19: ...19 This page intentionally left blank...

Страница 20: ...20 IFU SC3 001US REV 02...

Отзывы: