Note:

 After any of these adjustments a functionality 

check must be carried out (See section 11). 
 
 

8.

 

Footplate operation & adjustment 
 

 

The Configura Advance chair is supplied with a 
removeable angle adjustable footplate.  The footplate 
has 3 angle settings: 90°, 105°and 120°. 
The height adjustment system of the footplate is 
separate from the footplate removal 

 for details on 

how to adjust the footplate height refer to section 9 of 
this manual.  The footplate may be fitted, removed 
and adjusted with the patient in the chair.  Ensure the 
patient is sitting securely and cannot slide out. 
 

Operation 

 Fitting the Footplate 

1.

 

Ensure that the plungers that secure the footplate 
(Fig 21, Items 1&2) are in the closed position (Fig 
23). 

2.

 

Take the footplate (Fig 24) and insert the two 
tapered spigots (Fig 24, Items 1&2) into the two 
sockets either side of the footplate height 
adjustment frame (Fig 21, Items 3&4).  Push the 
footplate into the tubes until the plungers latch 
with an audible click.  

3.

 

Ensure the plungers are latched by moving the 
footplate in and out slightly. 

 

 

 

 

 

Operation - Removing the Footplate 

1.

 

Locate the plungers that secure the footplate 
either side of the leg rest, at the bottom of the 
footplate height adjustment frame (Fig 21, Items 
1&2). 

2.

 

Turn the levers on the two plungers towards the 
front of the chair until they latch. The lever arm 
plungers are now locked in the open position.  
(Fig 22). 

3.

 

Pull the footplate out. 

 
 
 
 
 
 
 
 

Figure 20 

 Actuator mounting 

WARNING 

Always engage the castor brakes when the 
chair is stationery or left unattended. 
Ensure the patient is in a stable sitting position 
while the footplate is adjusted, fitted or 
removed. 

Figure 21 

Figure 22 

 Plunger Open 

Figure 23 

 Plunger Closed 

Figure 24 

Low seat 

height position 

High seat 

height position 

Содержание Configura Advance SC3

Страница 1: ...Configura Advance Chair Instruction Manual IFU SC3 001US Rev 02 March 2021...

Страница 2: ...fied as a Class 1 Medical Device in accordance with the Medical Device Regulation 2017 745 Notice to User If a serious incident occurs in relation to this medical device affecting the user or the pati...

Страница 3: ...pump cable could lead to dangerous electrical hazards if squeezed or crushed between moving parts The chair must not be used if there is any visible damage to any cables 12 Do not move the chair when...

Страница 4: ...tended for use in primary healthcare environments where supervision can be provided if necessary for example long term residential care nursing homes or convalescent units The Configura Advance chair...

Страница 5: ...1 050 Dropdown sidepieces and covers set of 2 DRPARM 0 SC1 250 Dropdown armrests 400 550mm 16 22 set of 2 DRPARM 0 SC1 050 Dropdown armrests 350 500mm 14 20 set of 2 DRPARM 0 SC1 150 Dropdown armrests...

Страница 6: ...with the clevis at the back of the main frame Fig 3 Item 1 Ensure the Tilt in Space operating cable Fig 3 Item 4 is routed between the backrest and the backrest actuator assembly 6 Insert the clevis p...

Страница 7: ...e other side 10 Secure the backrest cushions to the backrest with the Velcro flaps Each cushion is attached to the backrest with the Velcro flap at the top of the cushion The top cushion has a double...

Страница 8: ...te Height range The footplate height can be adjusted from 330mm 13 to 533mm 21 The height is adjusted using two plungers 5 Chair Seat Height range The Advance chair frame can be adjusted to give seat...

Страница 9: ...h the other hand The depth setting can be read off the scale on the back part of the seat frame Fig 14 Item 1 Figure 13 5 Release the depth adjustment lever and move the backrest in and out slightly t...

Страница 10: ...footplate See section 8 and place the chair carefully on its front 4 Using a 13mm spanner and 5mm Allen key undo and remove the Nylock nut and shoulder bolt Fig 19 Item 1 and Item 2 Keep the bolt was...

Страница 11: ...ig 24 Items 1 2 into the two sockets either side of the footplate height adjustment frame Fig 21 Items 3 4 Push the footplate into the tubes until the plungers latch with an audible click 3 Ensure the...

Страница 12: ...can be used to restrict the use of the backrest angle adjustment 1 Tilt in Space The tilt in space position can adjust the seat angle from 7 degrees to 37 degrees from the horizontal position see pag...

Страница 13: ...e 7 If locking out the backrest is required return the locking lever to the locked position If the locking lever is too loose or too tight it may be necessary to rotate the whole lever clockwise or an...

Страница 14: ...to press Care must be taken to ensure the castor brakes are always locked when the chair is in use being assembled or dismantled so that the chair does not move accidentally Individually braked 125mm...

Страница 15: ...r can be braked at the same time using the foot lever 1 To lock the central brake press the rear red part of the foot lever down until the brake locks Figure 31 shows the brake locked 2 To unlock the...

Страница 16: ...s profiled headrest external laterals 3 Clean as described in the Cleaning Information 13 Troubleshooting Problem Possible solution s Backrest will not recline manual version Ensure the backrest locko...

Страница 17: ...tplate securely attached 12 Are the plungers securing the sidepieces locked and secure 13 Is there any risk of entrapment or patient injury 14 Are any accessory electrical cables pinched crushed or da...

Страница 18: ...t IPX4 Degree of protection against liquid ingress Do not dispose of in household waste Degree of protection against electric shock Class II Double Insulated For indoor use only Medical Device in acco...

Страница 19: ...19 This page intentionally left blank...

Страница 20: ...20 IFU SC3 001US REV 02...

Отзывы: