Gu id a n ce a n d m a n u f a ct u r e r ’s d e cla ra t io n – e le ct r o m a g n e t ic im m u n it y
Conduct ed RF
im m unit y
I EC 61000- 4 - 6
3 Vrm s
150 k Hz t o 80 MHz.
80 % AM at 1 k Hz
6 Vrm s in I SM bands
bet ween 150 k Hz and 80
MHz ( see not e)
3 Vrm s
150 k Hz t o 80 MHz.
80 % AM at 1 k Hz
6 Vrm s in I SM bands
bet ween 150 k Hz and 80
MHz ( see not e)
WARNI NG:
Port able RF com m unicat ions
equipm ent ( including peripherals such
as ant enna cables and ex t ernal
ant ennas) should be used no closer
t han 30 cm ( 12 inches) t o any part of
t he CADScor sy st em ACS- 1401,
including cables specified by t he
m anufact urer. Ot herwise, degradat ion
of t he perform ance of t his equipm ent
could result .
Radiat ed RF
im m unit y
I EC 61000- 4 - 3
3 V/ m
80 MHz t o 2.7 GHz
80 % AM at 1 k Hz
Plus addit ional t est lev els
according t o t he below
t able.
3 V/ m
80 MHz t o 2.7 GHz
80 % AM at 1 k Hz
Plus, addit ional t est lev els
according t o t he below
t able.
N O T E:
The I SM ( indust rial, scient ific and m edical) bands bet ween 0 .15 MHz and 80 MHz are 6.765 MHz t o
6.795 MHz; 13 .553 MHz t o 13.567 MHz; 26.95 7 MHz t o 27.2 83 MHz; and 40.66 MHz t o 4 0.70 MHz.