![acarix CADScor System Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/acarix/cadscor-system/cadscor-system_user-manual_3599219080.webp)
80
Er r or m essag e
D isp la y
r e fe r e n ce
Possib le ca use
Corr e ct ion a ct ion
# 4
Qualificat ion expired
REF: 300
The prev ious qualificat ion
has ex pired.
Qualify the sensor in t he Dock ing st at ion.
# 5
Qualificat ion failed
REF: 301
I ncorrect Dock ing stat ion.
Place sensor in hom e Dock ing st at ion.
REF: 303
Leak between the sensor
head and the Dock ing
st at ion sensor t ester.
Error in t he Dock ing st at ion.
Error in t he sensor.
Place t he sensor head firm ly int o t he
sensor test er and observe t hat red
indicat or ring on t he sensor head is not
v isible and ret ry qualificat ion.
Cont act local dist ribut or/ Acarix if sensor
will not qualify.
# 6
Am bient noise t oo
high
REF: 203
REF: 204
The acoust ic noise level in
t he exam inat ion room was
t oo high, abov e 65 dB SPL.
Close doors and windows to shield from
acoust ic noise com ing from outside
recording room .
Keep silent during the recording.
Turn off running nearby fans, radios or
sim ilar.
Change exam inat ion room locat ion.
Ot her ex am ples of noise:
Const ruct ion/ building m achinery , road &
rail t raffic noise, em ergency helicopter
serv ice and sim ilar.