14
5.11
I nt ended condit ions of use
The CADScor
®
Syst em is int ended t o be
used at m edical clinics and hospit als, at
room t em perat ure and below 2500 m et er
above sea level.
The pat ient shall be lying on t he back on
an exam inat ion bed ( supine posit ion)
during ident ificat ion of t he fourt h left I nt er
Cost al (I C4- L) space and recording of heart
sounds. Mark t he I C4- L posit ion aft er
locat ion, approxim at ely t wo cent im et res of
t he st ernum bone for placem ent of pat ch.
The CADScor
®
Sensor m ust only be used
t oget her wit h a CADScor
®
Pat ch. Pat ches
are single-use only and should not be
m oved on t he body aft er first adherence. I f
a pat ch has been m isplaced a new pat ch
m ust be used. The applied t im e of t he
CADScor
®
Sensor (including pat ch) is
norm ally less t han 15 m inut es.
Use t he CADScor
®
Syst em below 2500
m et ers.
Train t o ident ify left I C4 locat ion. Wrong
posit ion m ay influence result .
Do not at t em pt t o re-use a
CADScor
®
Pat ch. The pat ch is single use
only for hygienic reasons.