8.12
Short diast ole CAD- score
I n a few cases a st andard CAD- score
cannot be calculat ed based on a m inim um
diast olic period lengt h, however an
est im at ed CAD- score will be shown. The
short diast ole CAD- score is indicat ed by an
ast erisk ( * ) in front of t he CAD- score.
When a short diast ole- based CAD- score is
present ed, a possibilit y t o redo t he
recording is offered. Try t o relax (lower t he
heart rat e) t he pat ient furt her before
at t em pt ing a new recording. A m axim um
of four recording at t em pt s are allowed per
pat ch. I t is not recom m ended t o pursue
recording aft er four at t em pt s on t he sam e
pat ient .
A short diast ole- based CAD- score is not as
precise as t he st andard CAD- score, which
is why prom pt ing for a new recording is
encouraged ( REF: 128) .