ABUS TECHNOLOGIES INC.

 

 

12 

TagTemp 

By date/time

: Logging is stopped on user-defined date and time 

Interval

: Defines the interval between readings: hour, minutes and seconds. When the logging mode is 

set to the 

instantaneous

 the minimum interval is 1 second. For 

maximum

 and 

minimum

 values, the 

minimum interval is 10 seconds in both models. 

Misc

:  Shows  the  Logging  Mode  parameter  that  defines  how  measured  value  will  be  logged.  Options 

are:

 

Instantaneous

:  The  value  logged  will  be  exactly  the  value  measured  at  each  interval. 

Measurement  is  taken  at  the  end  of  the  defined  interval.  The  minimum  interval  between 
measurements is 1 second. 

Minimum

:  The  value  logged  will  be  the  minimum  value  found  within  ten  consecutive 

measurements  taken  within  an  interval.  The  minimum  interval  between  measurements  is  10 
second. 

Maximum

:  The  value  logged  will  be  the  maximum  value  found  within  ten  consecutive 

measurements  taken  within  an  interval.  The  minimum  interval  between  measurements  is  10 
second.

 

Alarms

: Enables an alarm that is triggered according to user-defined parameters. Values shown in the 

temperature unit selected. 

 

PDA Alarm Settings screen 

Clocks

:  Provides  access  to  Logger  and PDA  clocks.  When a new  configuration is  sent  to  the  logger, 

clocks are updated. 

Offset

:  allows  fine  offset  adjustment  on the measured value.  The 

Reset

  button  clears  changes  made 

and the logged values are the measured values again. 

 

Offset adjust screen 

After  configuring  clocks  in  the 

Settings

  screen,  click 

Apply

  to  send  this  configuration  to  the  Logger, 

returning to the 

Monitoring

 screen. 

Sending a new configuration implies the 

exclusion

 of all data collected present in the logger memory.

 

Содержание TAGTEMP

Страница 1: ...TAGTEMP Data Logger User Manual ABUS TECHNOLOGIES INC...

Страница 2: ...sonal injury or death Therefore comply fully with all Warning and Caution notices Information in this manual is intended only to assist our customers in the efficient operation of our equipment Use of...

Страница 3: ...t II 3 Optic Interface IR Link3 4 Palmtop User 5 5 6 6 7 4 Configuration 1 General information Field 2 Acquisition Field 3 Start Loggings Field 4 Stop Loggings Field 5 Channels Field 6 Palmtop User 8...

Страница 4: ...ograms 2 PRESENTATION 2 1 Technical Parameters Measuring Range Temperature 20 C 70 C Accuracy 0 5 C 25 C 2 C max along the entire measured range Note A measurement error can be fixed using the OFFSET...

Страница 5: ...OG While in stand by not logging or after a series of measurements it flashes once at every four seconds During logging it flashes twice at every four seconds Alarm Indicator AL Warns the user as to a...

Страница 6: ...nding wand IR LINK3 to it Click on the menu Port Clicking on the menu Port all free communication ports available in the computer will be listed usually COM2 once the mouse is frequently connected at...

Страница 7: ...ality of the LogChart II is available for the PDA Palm running the LogChart PalmOS software The program is delivered with the logger The stalled in the Palmtop through a HotSync process data synchroni...

Страница 8: ...start the communication between the logger and the software the Parameters Configuration window is then displayed as shown in figure Configuration Window below showing the current configuration and in...

Страница 9: ...e and time Via Start Stop Button starts and interrupts readings by pressing the Start Stop button in the frontal part of the logger for two seconds See Figure 01 Setpoint acquisitions start when a tem...

Страница 10: ...n the PDA infrared wand to the logger communication window run the LogChart Palm OS software wake up the logger and press Search Device in the Recorded Data screen When the PDA finds the logger the Mo...

Страница 11: ...ption is selected a new box to define the stop logging time is displayed By Start Button Press the Start_Now button from the Monitoring screen to start logging The PDA must be turned towards the logge...

Страница 12: ...within an interval The minimum interval between measurements is 10 second Maximum The value logged will be the maximum value found within ten consecutive measurements taken within an interval The min...

Страница 13: ...g the LogChart II menu During data transference a status bar indicates remaining data to be transferred Data offloading time is proportional to the number of readings logged At the end of data transfe...

Страница 14: ...res and configurations and details about data acquired General Information window 5 1 3 Acquisitions Table Window Data acquired by one or both input channels user defined are displayed in engineering...

Страница 15: ...base and tap on Details Recorded Data Details screen shows information about the data base View Data shows in table format the logged values and the date and time they were performed Press Delete to...

Страница 16: ...ding to local legal rules or returned to ABUSTEK 7 After placing batteries back to the logger make sure the cover is firmly attached to the socket 6 2 Troubleshooting FAULT POSSIBLE CAUSE RESOLUTION L...

Страница 17: ...modify the instrument 9 All connections should be tightened properly 10 Power supply should be constant should not be fluctuating 8 WARRANTY ABUS provides the original purchaser of this instrument a...

Отзывы: