Liquid QC (Quality Control) Schedules
This feature is used to define a liquid quality control (QC) plan for the i
‑
STAT Alinity instruments. A
customized QC testing schedule can include: the cartridge types and liquid QC fluids to be used, the
cartridge types that are enabled by running the liquid QC, and the schedule for performing liquid QC. Up
to 3 liquid quality control schedules can be defined. The options for liquid QC schedules are shown here
once, since the options are identical for all three.
Note: Not all products are available in all regions. Check with your local representative for
availability in specific markets.
Frequency
Select one of the following:
Off
This is the default.
Daily
Run liquid QC every day.
Every___________
Blank is filled in with a day of the week selection
from the drop down list.
Every_____ ______ of
each month
The first blank is filled in with first, second, third or
fourth chosen from drop down list. The second
blank is filled in with day of the week chosen from
the drop down list. Example - Every 2nd Thursday
of the month.
Every n days starting
on mmddyy
Schedule liquid QC to run according to a time
interval and a start date. The time interval is a
number of days in the range 28 to 90. Select the
start date from the pop-up calendar.
Time
This option is enabled only when a value other
than Off has been specified for Frequency.
Testing due at HH:mm Time of day when liquid QC testing is due to be
done. Specify a value for hours and minutes in the
range 00:00 to 23:59. Default is 00:00 (midnight).
Grace period
Additional amount of time allowed before and
after testing due time to perform liquid QC testing
if testing is not done at the scheduled time. After
this amount of time elapses, patient testing is
disabled. Grace period also determines when the
alert message for scheduled QC will display on the
instrument.
n hours
Number of hours. 8 hours is default. Range is 0 to
255.
Apply Schedule to
Select which months to perform testing. Any or all
months of the year can be selected.
Cartridge QC Profile
If Frequency is specified, define at least one
Cartridge QC Profile.
198
i-STAT Alinity — System Operations Manual
Art: 746300-01 Rev. N
Rev. Date: 02-Nov-2022
Содержание i-STAT Alinity
Страница 1: ......
Страница 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Страница 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Страница 39: ......
Страница 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Страница 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Страница 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Страница 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Страница 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Страница 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Страница 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Страница 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Страница 124: ...Cartridge Box Information 86 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 125: ...Cartridge Pouch Information 87 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 126: ...Cartridge Portion Pack Information 88 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 147: ......
Страница 148: ...This page intentionally left blank i STAT Alinity System Operations Manual Art 746987 01 Rev A Rev Date 02 Oct 2016 ...
Страница 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Страница 167: ......
Страница 168: ...This page intentionally left blank i STAT Alinity System Operations Manual Art 746988 01 Rev A Rev Date 02 Oct 2016 ...
Страница 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Страница 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Страница 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Страница 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...