Hematocrit
On this tab, select settings for testing Hematocrit.
Hematocrit Setting (Calibration by
Laboratory Hematology Analyzer)
Specify the anticoagulant used to calculate hematocrit result. For
best agreement of i
‑
STAT Alinity and hematology analyzer
hematocrit results, the i
‑
STAT Alinity customization setting is
selected according to the calibration of the comparative
hematology analyzer (MH
‑
K
2
EDTA or MH
‑
K
3
EDTA). Since most
clinical hematology analyzers are calibrated by the
microhematocrit method using K
3
EDTA anticoagulant, the i
‑
STAT
Alinity System default customization is K
3
EDTA.
Select either:
K2EDTA
K3EDTA
The default is K3EDTA.
Apply CPB Protein Algorithm to
Hematocrit
The Apply CPB option is intended for use when samples are
collected from patients on cardiopulmonary bypass. The CPB
function adjusts hematocrit and hemoglobin results for the
dilutional effect of pump fluid during cardiopulmonary bypass
surgery. However, the facility may validate its use for other
patient populations known to have protein levels significantly
lower than the normal adult population.
Note: If an instrument customized as “CPB, always
apply” is used for patients who are not on the pump, it
is possible that the hematocrit results will be reported
falsely high. If an instrument customized as “CPB, never
apply” is used for patients who are on the pump or have
decreased levels of protein, the hematocrit results may
be reported falsely low.
An instrument that is customized to prompt for CPB requires the
operator to answer Yes or No. Yes indicates that the CPB
correction should be applied. No indicates that the CPB
correction should not be applied. Select one of the following:
CPB, Prompt
Prompts the operator to apply CPB correction when the cartridge
has a hematocrit sensor.
CPB, always apply
Apply CPB correction every time a cartridge with a hematocrit
sensor is used.
CPB, never apply
Do not apply CPB correction when running a cartridge with a
hematocrit sensor.
186
i-STAT Alinity — System Operations Manual
Art: 746300-01 Rev. N
Rev. Date: 02-Nov-2022
Содержание i-STAT Alinity
Страница 1: ......
Страница 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Страница 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Страница 39: ......
Страница 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Страница 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Страница 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Страница 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Страница 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Страница 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Страница 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Страница 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Страница 124: ...Cartridge Box Information 86 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 125: ...Cartridge Pouch Information 87 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 126: ...Cartridge Portion Pack Information 88 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Страница 147: ......
Страница 148: ...This page intentionally left blank i STAT Alinity System Operations Manual Art 746987 01 Rev A Rev Date 02 Oct 2016 ...
Страница 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Страница 167: ......
Страница 168: ...This page intentionally left blank i STAT Alinity System Operations Manual Art 746988 01 Rev A Rev Date 02 Oct 2016 ...
Страница 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Страница 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Страница 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Страница 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...