5
ENGLISH
Measurement Units
The unit of measurement set by the factory appears. You cannot change this
setting. Contact Customer Service for more information.
Press
to return to set beeper, or press and hold
to turn off the meter.
Preparing the Meter for Glucose Testing
Important Testing Notes
This display always appears when the meter is turned on.
You should check that your meter matches the example you
see here exactly every time your meter turns on.
Do not use the meter if the display check screen does not
exactly match the example.
If the screen does not match, the meter may show an incorrect
result. Please contact Customer Service.
Calibration
Calibrate the meter to:
•
Match the test strips being used
•
Obtain accurate results
Calibration is required:
•
When using the meter for the first time
•
When opening and using a new box of test
strips
IMPORTANT: This display means the
meter is not calibrated for glucose testing.