Blood Glucose Control
Solution Test
Control Solution Results
Compare the control solution result to the
range printed on the test strip instructions
for use . The result on your screen should be
in this range .
Control solution results are marked on
the results screen and in the Logbook
with a
symbol .
Note:
Repeat the control solution test if the
results are outside of the range printed on
the test strip instructions for use . Stop using
the built-in meter if the control solution
results are repeatedly outside of the printed
range . Contact Customer Service .
Ketone Control Solution Test
100
mg
dL
22:23
Control Solution Test
0
.4
mmol
L
22:23
Ketone Control
Solution Test
Charging the Reader
A fully charged Reader battery should last up to 4 days . Your battery
life may vary depending on your usage . A
Low Battery
message
accompanies your result when you have enough charge remaining for
about one day of use .
Plug the included USB cable into an electrical
outlet using the included power adapter . Then,
plug the other end of the USB cable into the USB
port on the Reader .
Notes:
•
You must charge the Reader when the battery
is low
to keep using the Reader .
•
To fully charge the battery, charge the Reader
for at least 3 hours .
•
Only use the USB cable and power adapter
included with the system .
•
Fully charge your Reader before storing it for
more than 3 months .
OK
Please charge the Reader to
continue using it.
Low Battery
Charging
CAUTION:
Be sure to select a location for
charging that allows the power adapter to be
easily unplugged .