PL-0370 Rev 00 / December 2019 / Page 2

2  SWITCHING TO BACKUP HARDWARE

Switch all components (Console, Motor, Flow Probe and cables) simultaneously and 

then perform troubleshooting on the non-functioning System when it is no longer 

being used for patient support. 
The steps to switch to a backup Console, Motor and Flow Probe set are as follows:

© 2019 Abbott. All Rights Reserved.

Continue clamping the tubing while lowering the RPM to zero. Turn the main 

Console’s power switch off. Turn on the backup System, and make sure that it 

is properly assembled with the backup Motor.
Unthread the Pump retaining screw on the Motor by turning the screw counter-

clockwise several revolutions until the screw tip is clear of the locking groove 

on the Pump.

Place the Pump in the backup Motor receptacle (the Pump will drop into place 

in one of 3 orientations).
Rotate the Pump 

counterclockwise

 until it stops.

Thread the Pump retaining screw to secure the Pump in place by turning the 

screw clockwise until it stops. Confirm that the retaining screw is visible in one 

of the notches on the side of the pump. If the retaining screw is not visible in a 

notch, loosen the retaining screw, remove the pump, remount, rotate the Pump 

counterclockwise

, and secure by advancing the retaining screw.

Pumping operation may now be re-established via the backup Console 

by increasing the RPM to about 1000 RPM. Gradually increase RPM while 

unclamping the Pump outflow tubing. Continue to gradually increase the RPM 

to achieve the desired flow. Pumping operation should be returned to the last 

operating condition of the Pump. Be sure to reset options and alarms to match 

those selected before hardware exchange.

CentriMag

TM 

Circulatory Support System 

Alarms and Alerts Quick Reference Guide

NOTE: 

To be available for emergency use, the backup Console, Motor and Flow 

Probe must be transported with and in proximity of the main Console and Motor, 

and plugged into AC power, with the main power switch OFF. The battery power for 

the backup Console must be periodically assessed by turning the main power switch 

ON, checking the status of the battery, and then turning the main power switch OFF. 

In addition, the expiration date and maintenance due date of the backup Console’s 

Battery, must be checked on a periodical basis.

For additional information, refer to the CentriMag Circulatory Support System 

Operation Manual.

1

3

4

Rotate the Pump body clockwise until the grooves in the Pump match 

the Motor. Lift the Pump from the receptacle.

5

6

7

8

If the Pump has been OFF for more than five minutes without ade

-

quate anticoagulation, or if there has been a Motor Overheating con

-

dition, it will be necessary to replace the Pump and circuit, including 

cannulae.

2

Clamp the return tubing (outlet side).

1

2

3

4

5

6

7

8