AFINION
™
2 User Manual
US | 21
Testing Procedures
Collecting a sample
• The patient sample material and control material to be used is specific for each Afinion Test.
• The length of the capillary in the sampling device, and thereby the sample volume, might also vary for the different Afinion Tests.
• The time from filling the capillary until analysing the test cartridge must be as short as possible.
• Do not use the test cartridge if dropped on the bench or floor after the sample has been collected.
Consult the package insert that comes with each Afinion Test Kit for assay specific information.
Examples:
1
2
3
Remove the sampling device from the test
cartridge.
Use the handle to keep the test cartridge steady
against the table and pull the sampling device
straight up.
Fill the capillary; hold the sampling device almost
horizontally and bring the tip of the capillary in
surface contact with the sample. Make sure that
the capillary fills completely. It is not possible to
overfill.
Do not wipe off the capillary.
Avoid air bubbles and excess sample on the
outside of the capillary.
Immediately and carefully replace the
sampling device into the test cartridge.
The time from filling the capillary until
analysing the test cartridge must be as
short as possible.
Analysing a patient/control sample
1
2
3
Touch
to enter the patient sample mode.
Touch
to enter the control mode.
The lid opens automatically.
Insert the test cartridge with the barcode label
facing left.
Be sure that the test cartridge is correctly placed in
the cartridge chamber.
A “C” in the upper left corner indicates that the
analyzer is in the control mode.
Close the lid manually.The analyzer will start
processing the test cartridge.
The processing time depends on the test
in use.
4
5
6
Touch
and enter the patient ID.
Touch
to confirm.
Touch
and enter the control ID or Afinion
Control Data.
Touch
to confirm.
Entering the patient ID, control ID or Afinion Control
Data will not interrupt the processing.
Record the result, then touch
to accept.
If a printer is connected, touch
to print
the result.
The lid opens automatically.
The result will be saved in the result records.
Remove the used test cartridge from the
cartridge chamber and discard it in a
suitable waste container.
Insert a new test cartridge or close the lid
manually.
Keep the lid closed to protect the cartridge
chamber when the analyzer is not in use.
Consult the package insert that comes with each Afinion Test Kit for assay specific information.