14 | US
AFINION
™
2 User Manual
Configuration
Operator configuration
The operator ID function is disabled as a default setting by the manufacturer.
Touch
in the Configuration menu to enter the Operator configuration menu.
Operator ID enable/disable
Touch
in Operator configuration menu to enable/disable operator ID.
Select
to disable the operator ID function.
Select
to enable operator ID. Any operator ID is accepted.
Select
to enable operator ID with verification.
· To enable this function at least one supervisor is required to be present in the operator list.
· When operator ID with verification is enabled, analyzer configuration will only be available to the supervisors.
· To log in, the operator ID entered is required to be present in the operator list. See “Operator list management”, page 14.
Touch
to accept and return to the Configuration menu.
Operator login expiration
Touch
in the Operator configuration menu to set automatic logout of the operator.
Enter the number of minutes before automatic logout of operator.
The operator will automatically be logged out after the configured number of minutes after ended test.
Touch
to confirm and return to previous view.
Operator list management
Touch
in Operator configuration menu to enter operator list.
1
Touch
to add new operator.
Touch desired operator ID and touch
to delete or
to edit the highlighted operator.
Copy operator list
It is possible to copy an existing operator lists between analyzers using a USB flash drive. Insert USB flash in the analyzer USB port.
Touch
to export operator list from instrument to USB flash. Move USB to the new analyzer and touch
to import operator list. Any
existing operator list on the analyzer will be deleted.
2
Enter new/edit operator ID
Enter new/edit operator ID and touch
to enter. Both letters and numbers can be entered (maximum 16 characters).
If a barcode reader is connected to the analyzer, a barcoded operator ID can be entered.
3
Configure the operator level
Select USER to configure user access.
Select SUPERVISOR to configure supervisor access.
Configure tests accessible:
Select the test accessible for this operator.
Touch
to return and edit the operator ID.
Touch
to accept and store new operator in the operator list. The operator list can store 1000 operator IDs.
Supervisors will be marked with * in the operator list. When analyzer is configured to Operator ID with verification, configuration of the
analyzer will only be available to the supervisors.