6 SCC-K
NO
2
/ NO Converter Operator’s Manual
42/23-52 EN Rev. 2
Chapter 1
Converter Installation and Start-Up
Instructions for Selecting a Location
Short Gas Paths
The converter should be installed as close as possible to the analyzer system in
order to avoid the re-oxidation of NO and NO
2
in long sample gas lines.
Adequate Air
Circulation
Provide for adequate natural air circulation around the cooler unit. Avoid heat
buildup.
Protection from
Adverse Conditions
Protect the converter from
•
Cold
•
Heat sources such as the sun, ovens and vats
•
Large temperature variations
•
Strong air currents
•
Accumulations of dust and dust infiltration
•
Corrosive atmospheres
•
Vibration
Environmental
Conditions
Ambient temperature range +10 to +50 °C.
Average annual relative humidity
≤
75 %, occasional and slight condensation
permitted.
Converter Unpacking
•
If there is shipping damage which points to improper handling file a damage
claim with the shipper (railway, mail or freight carrier) within seven days.
•
Make sure that none of the enclosed accessories are lost during unpacking.
•
Keep the shipping box and packaging material for future shipping needs.
Catalyst Cartridge
The SCC-K NO
2
/ NO converter is a completely pre-installed unit. The standard
catalyst cartridge supplied is already installed.