ABB ProcessMaster FEM630 Скачать руководство пользователя страница 21

21

104 (Extended_Device_Status)

Data Type

Display

Bytes

Access

Device_Status

BYTE

Hex

1

Read Only

Mass_flowrate_exceeds_limits_0_0

BOOL

Decimal

Read Only

Volume_flowrate_exceeds_limits_0_1

BOOL

Decimal

Read Only

Simulation_is_on_0_2

BOOL

Decimal

Read Only

Flowrate_to_zero_0_3

BOOL

Decimal

Read Only

Maintenance_interval_is_reached_0_4

BOOL

Decimal

Read Only

All_totalizer_stop_0_5

BOOL

Decimal

Read Only

Totalizer_reset_0_6

BOOL

Decimal

Read Only

Display_value_is1600h_at_Qmax_0_7

BOOL

Decimal

1

Read Only

Device_not_calibrated_1_0

BOOL

Decimal

Read Only

Sensor_memory_defective_1_1

BOOL

Decimal

Read Only

NV_data_defect_Data_storage_1_2

BOOL

Decimal

Read Only

No_Frontend_Board_detected_1_3

BOOL

Decimal

Read Only

FEB_communication_error_1_4

BOOL

Decimal

Read Only

Incompatible_Frontend_Board_1_5

BOOL

Decimal

Read Only

NV_chips_defect_on_Motherboard_1_6

BOOL

Decimal

Read Only

Pulse_output_is_cut_off_1_7

BOOL

Decimal

1

Read Only

Current_output_3132_is_saturated_2_0

BOOL

Decimal

Read Only

CurrOut_V1V2_V3V4_saturated_2_1

BOOL

Decimal

Read Only

CurrOut_3132_com_error_2_2

BOOL

Decimal

Read Only

Option_Card_1_com_error_2_3

BOOL

Decimal

Read Only

Option_Card_2_com_error_2_4

BOOL

Decimal

Read Only

Safety_Alarm_CurrOut_3132_2_5

BOOL

Decimal

Read Only

CurrOut_3132_not_calibrated_2_6

BOOL

Decimal

Read Only

CurrOut_V1V2_not_calibrated_2_7

BOOL

Decimal

1

Read Only

CurrOut_V3V4_not_calibrated_3_0

BOOL

Decimal

Read Only

MB_voltages_outside_range_3_1

BOOL

Decimal

Read Only

An_alarm_is_simulated_3_2

BOOL

Decimal

Read Only

Communication_card_not_responding_3_3

BOOL

Decimal

Read Only

Reserved_0

BOOL

Decimal

Read Only

Coil_regulation_error_3_5

BOOL

Decimal

Read Only

Coil_wiring_Detection_3_6

BOOL

Decimal

Read Only

Coil Impedance measurement_3_7

BOOL

Decimal

1

Read Only

Electrode short circuit detection_4_0

BOOL

Decimal

Read Only

Electrode open circuit detection_4_1

BOOL

Decimal

Read Only

DC Feedback Regulation Error_4_2

BOOL

Decimal

Read Only

Monitoring Comm ADC and RX210_4_3

BOOL

Decimal

Read Only

Coil Isolation_4_4

BOOL

Decimal

Read Only

Gas bubble alarm_4_5

BOOL

Decimal

Read Only

Conductivity exceeds limits_4_6

BOOL

Decimal

Read Only

Sensor Temperature exceeds limits_4_7

BOOL

Decimal

1

Read Only

TFE alarm_5_0

BOOL

Decimal

Read Only

EPD alarm_5_1

BOOL

Decimal

Read Only

ADC Signal overrange_5_2

BOOL

Decimal

Read Only

SIL self check alarm_5_3

BOOL

Decimal

Read Only

Inhouse temperature exceeds limits_5_4

BOOL

Decimal

Read Only

Reserved3_5_5

BOOL

Decimal

Read Only

Reserved4_5_6

BOOL

Decimal

Read Only

Reserved5_5_7

BOOL

Decimal

1

Read Only

Reserved6

SINT

Hex

1

Read Only

Table 11  Ethernet/IP Alarms FEP63x / FEH63x / FEW53x

Содержание ProcessMaster FEM630

Страница 1: ...n 01 09 00 Measurement made easy Introduction This manual describes how to configure an Ethernet connection for the following electromagnetic flowmeters ProcessMaster Wafer FEM630 For more information...

Страница 2: ...EtherNet Cat5e cable 6 How to connect the shielding of the Ethernet cable 6 M12 plug option 6 RJ45 connector option 7 Wiring with different network topologies 8 Parameterization for start up 9 Etherne...

Страница 3: ...he case may be Operator shall establish and maintain any appropriate measures such as but not limited to the installation of firewalls application of authentication measures encryption of data install...

Страница 4: ...n Color code on the Ethernet card plug Port 1 X1 White orange Port 1 X2 Orange Port 1 X3 White green Port 1 X4 Green Port 2 X5 White blue Port 2 X6 Blue Port 2 X7 White brown Port 2 X8 Brown The color...

Страница 5: ...ts of the PoE card Port Pin Verion 1 Version 2 Version 3 Version 4 Port 1 X1 RD Pwr Spare RD Pwr Port 1 X2 RD Pwr Spare RD Pwr Port 1 X3 TD Pwr Spare TD Pwr Port 1 X4 TD Pwr Spare TD Pwr Port 2 X5 Spa...

Страница 6: ...res connecting to Port 1 2 These options allow for connection to different network topologies Topology 4 wires 4 wires 4 wires 8 wires 1 x M12 4 wires 2 x M12 4 wires 1 x M12 8 wires Star Y Y Y Ring o...

Страница 7: ...ach connecting to Port 1 and 2 Flowmeter equipped with 1 x RJ45 with 8 wires connecting to Port 1 and 2 These options allow for connection to different network topologies Topology 4 wires 4 wires 4 wi...

Страница 8: ...2 X5 Spare 1 white blue X6 Spare 1 blue X7 Spare 2 white brown X8 Spare 2 brown 1 4 Yes 1 X1 Recommendation use cable with 8 wires X2 X3 X4 1 8 Yes 1 X1 RD white orange X2 RD orange X3 TD white green...

Страница 9: ...st IP address is set c Set Webserver Access to Full d Set EtherNet IP Access to Full e Set Display Tag to Bus Address Communication Ethernet General Internet layer DHCP client enabled Host IP address...

Страница 10: ...st Red flashing 1 Hz Mod changing between R and Simple Error which can be fixed Red ON Mod showing R continously Major Error Non removable serious error OFF Mod showing continously No Power F Network...

Страница 11: ...s 5 Set Access to full Computer settings The IP address of the Ethernet adapter must be set up so that the subnet of the Flowmeter is accessible 1 Configure the Ethernet adapter for a fixed IP address...

Страница 12: ...es them internally 3 Wait for the device to verify the files 4 Make sure that a summary of the certificates shows on the webpage 5 If necessary upload new certificates or switch to a different certifi...

Страница 13: ...hrough the HMI menu Once a server is assigned an IP address the factory default IP address and hostname will be replaced by the ones assigned through the DHCP server To check the IP settings use the F...

Страница 14: ...Full Reset credentials Allows for resetting the webserver password Session timeout 5 min Timeout time closing the webserver session Default value 5 minutes Possible settings Min 1 Max 4294967000 Ethe...

Страница 15: ...tatus Example 100 MBits s P1 duplex status Example Full duplex P2 auto negotiation enabled disabled enabled P2 speed set P2 duplex set P2 speed status Example 100 MBits s P2 duplex status Example Full...

Страница 16: ...he Ethernet IP Interface TCP IP 0xF5 Attribute 5 DHCP DHCP Client enabled The device raises a DHCP request to obtain an initial IP address The DHCP Server detects this request and assigns an IP addres...

Страница 17: ...its_0_0 Volume_flowrate_exceeds_limits_0_1 Simulation_is_on_0_2 Flowrate_to_zero_0_3 Maintenance_interval_is_reached_0_4 All_totalizer_stop_0_5 Totalizer_reset_0_6 Display_value_is1600h_at_Qmax_0_7 De...

Страница 18: ...e SINT Hex 1 Read Only Table 6 Assembly Instance 101 Totalizer 102 Diag Values Data Type Display Bytes Access Conductivity REAL Float 4 Read Only Coil_Temperature REAL Float 4 Read Only Inhouse_Temper...

Страница 19: ...mal Read Only Reserved_0 BOOL Decimal Read Only Coil_regulation_error_3_5 BOOL Decimal Read Only Coil_wiring_Detection_3_6 BOOL Decimal Read Only Coil Impedance measurement_3_7 BOOL Decimal 1 Read Onl...

Страница 20: ...rm bits If an error occurs one or more corresponding warning and or alarm bits are set If an error disappears one or more corresponding warning and or alarm bits are reset This data type also includes...

Страница 21: ...ut_3132_2_5 BOOL Decimal Read Only CurrOut_3132_not_calibrated_2_6 BOOL Decimal Read Only CurrOut_V1V2_not_calibrated_2_7 BOOL Decimal 1 Read Only CurrOut_V3V4_not_calibrated_3_0 BOOL Decimal Read Onl...

Страница 22: ...Code Service Name 0x0E GAS Table 15 Instance Services Attr ID Name Access Rule Data Type 1 Vendor ID Get UINT 2 Device Type Get UINT 3 Product Code Get UINT 4 Revision Get STRUCT of 0 Major Revision G...

Страница 23: ...Attributes Assembly 0x04 The Assembly Object binds attributes of multiple objects that data can be sent or received from any object over a single connection Assembly objects can be used to bind input...

Страница 24: ...Type 1 Open Requests Get Set UINT 2 Open Format Rejects Get Set UINT 3 Open Resource Rejects Get Set UINT 4 Open Other Rejects Get Set UINT 5 Close Requests Get Set UINT 6 Close Format Rejects Get Se...

Страница 25: ...ifferent priorities or other delivery characteristics Standard QoS mechanisms include IEEE 802 1D Q Ethernet frame priority and Differentiated Services DiffServ in the TCP IP protocol suite The QoS ob...

Страница 26: ...ed EPATH 10 Port Routing Capabilities Get UDINT Table 46 Instance Attributes TCP IP Interface 0xF5 The TCP IP Interface object configures the TCP IP network interface of a device such as the IP addres...

Страница 27: ...T ArpPdu 0 ARRAY of 28 USINT 13 Encapsulation Inactivity Timeout Set UINT 16 Active TCP Connections Get UINT 17 Non CIP Encapsulation Messages s Get UDINT Table 50 Instance Attributes Ethernet Link 0x...

Страница 28: ...UDINT 0 Single Collisions 0 UDINT 0 Multiple Collisions 0 UDINT 0 SQE Test Errors 0 UDINT 0 Deferred Transmissions 0 UDINT 0 Late Collisions 0 UDINT 0 Excessive Collisions 0 UDINT 0 MAC Transmit Error...

Страница 29: ...per day ft3 d 1465 milliliter per second ml s 1407 milliliter per minute ml min 1411 liter per second l s 1406 liter per minute l min 1413 liter per hour l h 1414 liter per day l d 1446 hecto liter pe...

Страница 30: ...package is valid the webserver displayes the firmware version and checksum The checksum is an SHA 256 checksum allowing for package integrity check 5 To install the firmware on the Ethernet card selec...

Страница 31: ...y Version and checksum for the current Ethernet card firmware is Device firmware is not available for download from the ABB library Please refer to your local ABB Service Organization if a firmware up...

Страница 32: ...or possible lack of information in this document We reserve all rights in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or ut...

Отзывы: