background image

 

MR Series FT-Spectroradiometers  

22

desiccant material is satisfactory (by color.) The frequency of this schedule should be dependent upon the 
environment and for cases where the ambient humidity level is unchecked the frequency should be increased. 
Typically, the frequency to check the desiccant (and optical materials) should be on a monthly basis. If the 
desiccant is not checked periodically, it can become a source of moisture that may damage optical materials. 
As a side note, desiccant materials can usually be regenerated by heating the material to rid excess moisture. 
This can be accomplished by placing the desiccant in an oven and heated at 100 degrees Celsius for some period 
of time until the color comes back to the “dry” state. An alternate method can be to microwave the desiccant. A 
power level of 80% for approximately 10 minutes should regenerate the desiccant to its “dry” state, BUT make 
sure that the evaporated moisture can escape from the desiccant enclosure. Consult with the supplier of the 
desiccant if you are unsure in this respect.  

Heating  

By heating the optical elements above ambient temperature levels encountered, it can be ensured that the 
element remain above the temperature at which condensation may occur onto the surface of the material. [See 
Appendix I-4 for discussion of relative humidity.] This can be achieved by raising the temperature locally by 
using some appropriate device. If no other means are available, a small heat source such as a 25 Watt light bulb 
placed under and near enough to the component (both sides) that may be used. The temperature should be 
maintained at least 5 degrees Celsius over the ambient temperature to avoid condensation, 

BUT THIS WILL 

NOT 

ACCOUNT FOR RELATIVE HUMIDITY LEVELS REACHING LEVELS THAT MAY HARM THE OPTICAL MATERIAL

. Despite 

a 5 degree temperature difference, the relative humidity may approach levels much greater than the specified 
35% level. To ensure that the materials are preserved, a combination of heating slightly above ambient along 
with the use of desiccant material would be recommended with frequent observation. 

2.  Instrumentation 

 

The instrumentation also includes possible hygroscopic materials (windows and beamsplitters) that should be 
stored carefully. The same guidelines that apply to the optical elements also apply to the instrumentation, 
however, the optical elements may be permanently in place and cannot be removed for storage. The MR 
spectrometers possess a storage for desiccant material within the casting and should be regularly checked (see 
operating manual for further details.) If the relative humidity is expected to constantly, or frequently, be higher 
than 35%, then a dry oil free gas purge to the instrument housing should be applied. (Consult your ABB 
representative on this aspect.) Exposed hygroscopic materials on the instrumentation, such as a window on top 
plate or at side port location, can be protected by placing desiccant material near the exposed surface and to 
cover and seal the optical element and desiccant with a non-water-absorbing material. In order to maintain the 
temperature above ambient levels, it is recommended that the instrument remain “ON” during storage so the heat 
generated by its own operation maintain the temperature above ambient levels. 

iii.  Transport conditions 

 

At some time it may become necessary to transport instrumentation as well as the optical elements. The best and 
safest way this can be achieved is to place the optical materials in a well packed and sealed container using 
desiccant material (as discussed in Appendix I.3.2.1.) Exposed hygroscopic materials on the instrumentation, 
such as a window on top plate or at side port location, can be protected by placing desiccant material near the 
exposed surface and to cover and seal the optical element and desiccant with a non-water-absorbing material. 
Internal optical elements can be protected by checking that the desiccant material is satisfactory (see operating 
manual of the instrument for details on this aspect.)  

NEVER open cartons containing hygroscopic components in a receiving area, especially in cool weather. Move 
the cartons to a designated storage area (constant temperature) where the relative humidity is below 35%, and 
wait at least 24 hours to ensure a thermal equilibrium with its surroundings. 

d.  RELATIVE HUMIDITY 

 

Содержание MR Series

Страница 1: ...MR Series FT Spectroradiometers Start Up Procedure This manual contains Information on the unpacking and set up of the MR Series FT Spectroradiometers IMZ8947 Revision 3 1 March 2003 ...

Страница 2: ......

Страница 3: ...ifications of all equipment and software and contents of this document without obligation to notify any person or organization of such changes Every effort has been made to insure that the information contained in this document is current and accurate However no guarantee is given or implied that the document is error free or that the information is accurate ABB provides product support services t...

Страница 4: ......

Страница 5: ...HE CONFIGURATION FILE ACQUIRE CFG 16 4 2 1 Changing the instrument type 16 4 2 2 Running Research Acquire without instrument 16 4 2 3 External Trigger address setting 17 5 SHUTDOWN SEQUENCE AND SYSTEM PACKING 17 6 HANDLING AND CARE OF HYGROSCOPIC MATERIALS 19 6 1 INTRODUCTION 19 6 2 HYGROSCOPIC MATERIALS 19 6 3 HANDLING AND CARE 21 6 3 1 Handling of components 21 6 3 2 Storage conditions 21 6 3 2 ...

Страница 6: ......

Страница 7: ...nt up and running It also contains useful reference information on the set up of the instrument hardware and software and on packing the instrument in order to ship it This section also contains information on the handling of hygroscopic materials and should be studied by customers having ordered optional hygroscopic optics for their MR system ...

Страница 8: ......

Страница 9: ...G LN2 DETECTOR DEWARS THE MANUAL YOU ARE PRESENTLY READING TRIPOD TRIPOD Upon reception of the material check for any external damage to the carrying cases If any damage is observed please notify the courier and have this stated on the packing slip This is important for making a claim should the equipment have been damaged during shipment Please notify the After Sales Service Department at ABB Bef...

Страница 10: ...OPE DETECTOR DETECTOR MEDIUM ANGLE TELESCOPE WIDE ANGLE TELESCOPE CCD CAMERA MONITOR CCD CAMERA POWER SUPPLY CCD CAMERA 90 INPUT COLD SOURCE DATA CABLE SPARE FUNEL MANUALS Computer COMPUTER OR SPARE CABLES MANUALS AND KEY BOARD MOUSE MONITOR COMPUTER OR SPARE COMPUTER OR SPARE BACKUP SOFTWARE MR FT Spectroradiometer SPECTRORADIOMETER Accessories ...

Страница 11: ...MR100 In addition to the data connector the resolution setting knob the AC power inlet input power selection on off switch and status LED s can be found on this connector plate The POWER led is always on when the instrument is powered up The other LEDs are used to provide the instrument status during power and operation When the instrument is first turned on the SCAN and DATA LEDs comes on to indi...

Страница 12: ...ncy at which the DATA LED s flashes changes accordingly Please note that the data cable can be connected or disconnected at any time whether the unit is turned on or not Doing so will not harm the equipment MR200 connector plate showing fibre optics connector MR200 In addition to the data connector the AC power inlet input power selection on off switch and status display can be found on this conne...

Страница 13: ...ion must be clearly defined 7 fringe counter initialisation with ZPD location successful normal scanning ZPD position not same as previous two different consecutive ZPD positions ZPD position not same as previous two different consecutive ZPD positions 8 ZPD located once requires three consecutive identical ZPD positions ZPD located once requires three consecutive identical ZPD positions 9 fringe ...

Страница 14: ...iometers 8 MR100 The DSP board occupies two ISA slots in the host computer Only one external connector is available and is used for the connection of the DB 25 serial data cable The board is displayed on the next two figures ...

Страница 15: ...t computer Two connectors are available On is used for the connection of the fibre optic data cable The other DB 25 is an unused RS232 port This port could be used to receive RS232 signals coming from an IRIG B time code converter The board is displayed on the next two figures ...

Страница 16: ...of the DSP above for their locations on the DSP board You may also perform a visual inspection of the proper insertion of the ISA connector in the slot Should it look outside or not evenly inserted in the slot please push gently on the DSP board making sure that it is well inserted in the slots When the computer is not purchased from ABB the DSP board comes delivered as a separate item To install ...

Страница 17: ...he computer and make sure that it boots normally indicating no major hardware problem 12 At this point the computer is ready to be connected to the Spectroradiometer to perform a basic communication test c PUTTING THE BASIC SYSTEM TOGETHER An operational Spectroradiometer requires the following basic elements to be interconnected together The Spectroradiometer unit must be connected to the compute...

Страница 18: ...les to connect and the size of the connectors differs permitting only one location per cable This is not required for the MR200 as the connection is made automatically when the detector is installed in place This is done via a 50 pin connector installed on the output optics plate Make sure that cryogenic detectors are properly filled with Liquid Nitrogen 7 At this point the detector module s is ar...

Страница 19: ...e correct LED sequence is followed 13 Start the acquisition software from the computer please refer to the next section or the software manual and go into the alignment status screen where the interferogram and spectral data are displayed in real time 14 At this point the system is operational and data can be acquired ...

Страница 20: ...ll program will install the software into C ACQUIRE but this can be changed to install the software in any directory on the computer The install pro gram also needs to know what kind of instrument will be running with Research Acquire in order to correctly configure the software1 use No Instrument on computers that are only going to be used for post processing and not for data acquisition Once the...

Страница 21: ...Start Up Procedure 15 Figure 2 Communication initialisation screen Figure 3 Start up screen for Research Acquire ...

Страница 22: ...rument for example from MR100 to MR200 edit the parameter instrument type parameter and set it to the proper value MR100 1 and MR200 2 ii Running Research Acquire without instrument If you need to run Research Acquire without instrument edit the parameter acquisition board and set it to 1 This setting can be u seful if you have to process data after acquisition without the instrument plugged This ...

Страница 23: ...roradiometer OFF To prepare the spectroradiometer for shipment the following assemblies must be removed and properly placed in their respective carrying case Data cable Input telescope Detectors Cold Source Eye piece or CCD camera power supply and monitor IMPORTANT Turn the spectroradiometer off before removing the detectors Before moving the spectroradiometer into its carrying case make sure the ...

Страница 24: ......

Страница 25: ...ium bromide KBr potassium chloride KCl sodium chloride NaCl and many others While these materials provide relatively high transmittance of infrared radiation they are also susceptible to damage by interaction with atmospheric moisture The damage to these materials caused by moisture will be to introduce absorption features water bands and or to degrade the surface quality of polished surfaces that...

Страница 26: ...F2 TlBr TlCl KRS 5 KRS 6 0 6 0 5 0 4 0 3 0 2 0 1 0 Figure 5 1 Solubility of various optical materials Solubility of optical materials in H2O Optical material Solubility g 100g H2O Temperature of 25 degrees Celsius CsBr CsI KBr KCl KF KI NaBr NaCl NaI 190 180 170 160 150 140 130 120 110 100 90 80 70 60 50 40 30 20 10 0 Figure 5 2 Solubility of various alkali halide optical materials ...

Страница 27: ... may collect during handling Note that unused solvents should be discarded as they have a high affinity to absorb water vapor from the air and also materials that are likely to absorb moisture tissues tape clothing etc should be removed from close proximity of the optical materials ii Storage conditions 1 Optical materials When the optical materials that are hygroscopic are not in use it is recomm...

Страница 28: ...recommended with frequent observation 2 Instrumentation The instrumentation also includes possible hygroscopic materials windows and beamsplitters that should be stored carefully The same guidelines that apply to the optical elements also apply to the instrumentation however the optical elements may be permanently in place and cannot be removed for storage The MR spectrometers possess a storage fo...

Страница 29: ...perature of objects cools below the temperature of the air resulting in a dew forming on car windows plants and so forth With this in mind when the relative humidity is requested to be below a certain level it is important to note this temperature dependency and is the reason why the optical materials are requested to remain above ambient temperature levels to ensure that the material remain above...

Страница 30: ...of optical material failure the cause has been due to fogging of the component due to exposure to moisture Such problems can be eliminated by following the guidelines as laid out in this document The rule of thumb to be followed is that the relative humidity should remain below 35 at all times to preserve the quality of the hygroscopic optical materials Care should be taken in the storage handling...

Отзывы: